Name
Access
SequenceRoi
Gainn
n = 1 ~ 10
SequenceRoi
Hbinningn
n = 1 ~ 10
SequenceRoi
Vbinningn
n = 1 ~ 10
SequenceRoi
LutEnablen
n = 1 ~ 10
SequenceRoi
BlackLeveln
n = 1 ~ 10
SequenceRoi
GainRedn
n = 1 ~ 10
SequenceRoi
GainBluen
n = 1 ~ 10
SequenceRoi
Indexread
SequenceReset
Short
Min ~ Max
ASCII
R/W
SQGAn 100 ~1600
1: Hbinning =
OFF
R/W
SQHBn
2: Hbinning =
x2
1: Vbinning =
OFF
R/W
SQVBn
2: Vbinning =
x2
0: Off
R/W
SQLUTn
1: On
R/W
SQBLn
- 256 ~ 255
R/W
SQPGRn -4533 ~ 17713
R/W
SQPGBn -4533 ~ 17713
0: Index0
1: Index1
2: Index2
3: Index3
4: Index4
R/O
SQIDX
5: Index5
6: Index6
7: Index7
8: Index8
9: Index9
W/O
SQRST 0
Default
SQGA n=[Param.]<CR><LF>
SQGA n?<CR><LF>
100
Set the gain of Sequence Roi Index n.
SQHB n=[Param.]<CR><LF>
SQHB n?<CR><LF>
1
Set the horizontal binning of Sequence Roi Index n.
SQVB n=[Param.]<CR><LF>
SQVB n?<CR><LF>
1
Set the vertical binning of Sequence Roi Index n.
SQLUT n=[Param.]<CR><LF>
SQLUT n?<CR><LF>
0
Enable/disable the LUT setting for Sequence Roi Index n.
SQBL n=[Param.]<CR><LF>
SQBL n?<CR><LF>
0
Set the black level of Sequence Roi Index n.
SQPGR n=[Param.]<CR><LF>
SQPGR n?<CR><LF>
0
Set the Gain Red level of Sequence Roi Index n.
SQPGB n=[Param.]<CR><LF>
SQPGB n?<CR><LF>
0
Set the Gain Blue level of Sequence Roi Index n.
SQIDX?<CR><LF>
-
Dispaly the index number of the current Command Sequence
Index.
SQRST=0<CR><LF>
0
Reset the current index number for Trigger Sequence Mode
and Command Sequence Mode to "Index 1".