5920 UHP Moisture Analyzer Specifications

Specification
Range
Limit of Detection
Accuracy
Response Time
Sample Inlet Pressure
Requirement
Exhaust Pressure
Sample Flow Requirement Less than 0.5 SLPM
Inlet Gas Temperature
Outputs
Alarms/Alerts
Environmental Conditions
Utility Requirements
Instrument Air
Mounting
Dimensions
Net Weight
5-2 | 5920 UHP Moisture Analyzer
Description
Calibrated range of 0–150 PPBV. Trend indication to 1000 PPBV.
1 PPBV nominal (RMS Noise ≤0.3 PPBV)
±1 PPBV or ±10 % of value, whichever is greater
Typical 80 % of a 25 PPBV step change in less than 10 minutes
138–345 kPa (20–50 PSIG)
Specified performance obtained when inlet pressure is
maintained to within ±17 kPa (±2.5 PSIG).
Atmospheric
0 °C to 100 °C (32 °F to 212 °F)
Optimal results are obtained when inlet gas temperature is
maintained at 60 °C (140 °F).
4-line x 20-character LCD display
One (1) 4–20 mA DC analog output
RS-232 / RS-485 serial port
Three (3) independent contact closures of 30 VAC or 60 VDC
maximum, 10 VA or 1 A. Resistive for System alarm, moisture
Concentration alarm, and Data Valid. All are fail-safe by default.
Alarms are available on RS-485 interface.
10 °C to 30 °C (50 °F to 86 °F)
(Optimal results when ambient temperature is maintained at a
constant value ±5 °C)
90 % RH maximum, non-condensing
IEC Pollution Degree 2
Maximum elevation 2,000 meters (6560 feet)
IEC Installation Category II
Indoor use only, IEC 529; IP40
100–127 VAC, 47–63 Hz, 185 W
or
200–240 VAC, 47–63 Hz, 185 W (switch selectable)
550–690 kPa (80–100 PSIG)
-40 °C dew point
5 micron filter recommended
19" rack
Width:
48 cm
(19")
Height:
17.7 cm (7")
Depth:
50.9 cm (20")
15.9 kg (35 lb)