Ametek 5920 UHP User Manual page 86

Moisture analyzer for ultra-high purity gases
Table of Contents

Advertisement

5. Remove the female nut from the valve on the MFM to disengage the tub-
6. Once you have detached the Reference Valve solenoid bracket and the
7. Remove the two (2) screws that hold the MFM mounting bracket to the
8. To remove the MFM, remove the two (2) hex screws that hold the MFM to
9. Disconnect the cable assembly from the MFM.
4-6 | 5920 UHP Moisture Analyzer
ing.
tubing, you should be able to lift the solenoid bracket and attached Refer-
ence Purifier solenoid and tubing out of the enclosure.
mounting plate and lift the MFM, with the mounting bracket and remain-
ing solenoids attached, from the enclosure.
the mounting bracket. Gently pull the MFM away from the bracket (Figure
4-3) by sliding it out from the end of the mounting bracket.
PN 592047901 Rev YC