The Mass Flow Meter (MFM) (AMETEK Part No. 271189001) is attached to the
lower section of the MFM mounting bracket. See Figures 4-1 and 4-2.
To remove the Mass Flow Meter from the 5920 Analyzer:
1. After power has been disconnected, remove the top cover and sample
2. Disconnect the four (4) fittings on the solenoids labeled "SAM" and "REF" to
3. Disconnect the single solenoid for the Reference Purifier labeled "SAM 1"
4. Remove the two (2) hex screws that hold SAM 1 Reference Valve solenoid
Figure 4-2.
Location of MFM, mounting
bracket and valves.
PN 592047901 Rev YC
oven cover from the analyzer and locate the MFM mounting bracket. The
solenoid valve mounting brackets are attached to the top of the MFM
bracket.
free them from the associated tubing. The solenoid valves are still attached
to the solenoid mounting bracket and the MFM mounting bracket.
by loosening the female nut that attaches the tubing to the purifier on one
end and the nut attached to the MFM tubing on the other end.
bracket to the top of the MFM mounting bracket.