Operating Instructions
Series 505-1320
Level Measurement
Installation and
Need More Instruction Books ?
Go to www.drexelbrook.com
Look under this specific product
Download Instruction Manual
Leader in

Advertisement

Table of Contents
loading

Summary of Contents for Ametek Drexelbrook 505-1320 Series

  • Page 1 Outside North America + 215-674-1234 Installation and Operating Instructions Series 505-1320 Continuous Ultrasonic Transmitter using 405-1302 Electronics + 215-674-1234 Outside North America Need More Instruction Books ? 1-800-553-9092 US and Canada Go to www.drexelbrook.com www.drexelbrook.com Look under this specific product Download Instruction Manual drexelbrook.service@ametek.com...
  • Page 2 AMETEK Drexelbrook makes no warranty of any kind with regard to the material contained in this manual, including, but not limited to, implied warranties or fitness for a particular purpose. Drexelbrook shall not be liable for errors contained herein or for incidental or consequential damages in connection with the performance or use of material.
  • Page 3 EDO # 12-02-243 505-1302-LM Series 505-1320 Continuous Ultrasonic Transmitter using 405-1302 Electronics 205 Keith Valley Road Horsham, PA 19044 US Sales 1-800-553-9092 24 Hour Service 1-800-527-6297 International + 215-674-1234 DREXELBROOK + 215-674-2731 E-mail drexelbrook.info@ametek.com An ISO 9001 Certified Company www.drexelbrook.com...
  • Page 5: Table Of Contents

    EDO# 12-02-243 505-1302-LM Table of Contents SECTION 1 - INTRODUCTION ..................1 Product Description ....................1 Models Available ....................1 Models Available ....................1 Types of Flumes and Weirs ..................2 SECTION 2 - INSTALLATION ..................4 Unpacking ......................4 Mounting Instrument ....................
  • Page 7: Section 1 - Introduction

    505-1320 Series Ultrasonic Open Channel Flowmeter SECTION 1 INTRODUCTION The Drexelbrook Series 505-1302 Open Channel Flowme- Product Description ter accurately measures continuous open channel flow up to a range of 99.9 inches, using ultrasonic technology. The flowrate output is a 4-20 mA dc signal. Ultrasonic sensor is made of CPVC, Teflon and 316 SS or Teflon Sealtyte for compatibility with wide range of process materials.
  • Page 8: Types Of Flumes And Weirs

    505-1320 Series Ultrasonic Open Channel Flowmeter 1.4 Types of Flumes& Weirs Figures 1-1 through 1-8 show the 505-1302 Ultrasonic Open Channel Flowmeter mounted in different types of flumes and weirs. NOTE NOTICE The open channel flowme- ter location is a function Figure 1-1 of the primary measuring Parshall Flume...
  • Page 9 505-1320 Series Ultrasonic Open Channel Flowmeter 1.4 Types of Flumes& Weirs (continued) NOTE NOTICE The open channel flowmeter location is a function of the primary measuring device. Consult the flume/weir Figure 1-4 manufacturer for the proper V-Notch Weir sensor location. Figure 1-5 Palmer Bowlus Flume Figure 1-6...
  • Page 10: Section 2 - Installation

    505-1320 Series Ultrasonic Open Channel Flowmeter SECTION 2 INSTALLATION Carefully remove the contents of the shipping carton and Unpacking check each item against the packing list before destroying any packing material. If there is any shortage or damage, report it to the factory immediately. Refer to Figures 2-1 and 2-2 for mounting dimensions of Mounting Instrument the integral and remote instruments, respectively.
  • Page 11 505-1320 Series Ultrasonic Open Channel Flowmeter Mounting Instrument (continued) Figure 2-1 Mounting Dimensions Integral Unit...
  • Page 12 505-1320 Series Ultrasonic Open Channel Flowmeter Mounting Instrument (continued) Figure 2-2 Mounting Dimensions Remote Unit...
  • Page 13: Wiring

    505-1320 Series Ultrasonic Open Channel Flowmeter Wiring CAUTION If the Series 505-1302 is located in a hazardous environ- ment, do not open enclosure cover or make/break any electrical connections without first disconnecting electri- cal power at the source. Ensure that wiring, electrical fittings and conduit connections conform to electrical codes for specific location and environment.
  • Page 14 505-1320 Series Ultrasonic Open Channel Flowmeter Wiring (continued) Figure 2-3 Wiring Diagram of Line-Powered Transmitter...
  • Page 15 505-1320 Series Ultrasonic Open Channel Flowmeter Wiring (continued) Figure 2-5 Wiring Diagram of 24 Vdc Transmitter...
  • Page 16: Section 3 - Operation

    505-1320 Series Ultrasonic Open Channel Flowmeter SECTION 3 OPERATION Modes of Operation When using the 505-1302 Open Channel Flowmeter, there are two modes of operation: flow or distance mode. • • • • • In the flow (or level) mode, the current output (4-20 mA) increases as the liquid level rises.
  • Page 17 505-1320 Series Ultrasonic Open Channel Flowmeter Indicators & Controls —Alarm Relays Outputs for Near Zone and Lost (continued) Echo are provided to activate external relays or alarms. These are separate relay packages purchased as an option. —Step Gain Switch Changes the gain of the ultra- sonic transmitter to lessen any effect due to noise or obstruction.
  • Page 18: Time Delay

    505-1320 Series Ultrasonic Open Channel Flowmeter Time Delay Increasing time delay to either 15 or 45 seconds will smooth out a jumpy output signal caused by wave action or turbu- lence. Step Gain Switch The step gain switch can be used to decrease the acoustic signal and avoid noise interference.
  • Page 19: Section 4 - Calibration

    505-1320 Series Ultrasonic Open Channel Flowmeter SECTION 4 CALIBRATION The 505-1302 Series transmitter must be set for the type and size 4.1 Introduction of the primary device and the actual zero and span values for the specific application. This procedure is outlined in Sections 4.1 through 4.3.
  • Page 20 505-1320 Series Ultrasonic Open Channel Flowmeter Figure 4-1 Flow and Relay Switches...
  • Page 21: Standard Calibration

    505-1320 Series Ultrasonic Open Channel Flowmeter Standard Calibration Use this procedure to set the Zero and Span values. a. With the power off, set the Zero switches in inches (to the nearest 0.1 inch) to equal the distance from the transducer face to the level that represents 0% flowrate (4mA).
  • Page 22: Secondary Calibration

    505-1320 Series Ultrasonic Open Channel Flowmeter Secondary Calibration CAUTION Secondary calibration procedure permanently overwrites factory calibration. a. Perform the standard calibration per Section 4.3. b. Record the Zero switch settings. c. Measure the distance from the transducer face to the liquid level as accurately as possible (to the nearest 0.1 inch).
  • Page 23: Setpoint Calibration

    505-1320 Series Ultrasonic Open Channel Flowmeter Setpoint Calibration a. Refer to Figure 4-1 for location of relay switches. (only for those units equipped with optional b. Alarm points are set directly in tenths of inches 24 Vdc output) from zero point. As an example, an alarm point could be set 6 inches from the zero point or minimum level.
  • Page 24: Accessory Packages

    505-1320 Series Ultrasonic Open Channel Flowmeter Accessory Several accessories are available for the 505-1302 Ultrasonic Open Channel Flow instrument. Packages As an option, a 401-600 external relay package is available. 4.6.1 401-600 Series Relay Package This package provides a remote-mounted relay outputs for Alarm 1, Alarm 2, Lost Echo and Near Zone.
  • Page 25 505-1320 Series Ultrasonic Open Channel Flowmeter 4.6.1 401-600 Series Relay Package (continued)
  • Page 26 505-1320 Series Ultrasonic Open Channel Flowmeter 4.6.1 401-600 Series Relay Package (continued)
  • Page 27 505-1320 Series Ultrasonic Open Channel Flowmeter 4.6.1 401-600 Series Relay Package (continued) Figure 4-8 Relay Package NEMA 4X Housing with Intrinsic Safety Barriers...
  • Page 28 505-1320 Series Ultrasonic Open Channel Flowmeter 4.6.1 401-600 Series Relay Package (continued) Figure 4-9 Relay Package with Digital Loop Wiring...
  • Page 29: Sta-Dry™ Option (Modification 9128)

    505-1320 Series Ultrasonic Open Channel Flowmeter 4.6.2 Sta-Dry™ Option In some applications, condensation or ice can build up on the transducer face causing a reduction or loss of the (Modification 9128) acoustic signal. The Sta-Dry option maintains the temperature of the transducer face at a level that prevents condensation or ice from forming.
  • Page 30: Universal Process Meter (401-90X-2)

    505-1320 Series Ultrasonic Open Channel Flowmeter 4.6.3 Universal Process Meter As an option, the universal process meter is available (401-90X-2) with flow rate, totalizer and batch control capabilities. Before wiring the the universal process meter in the loop, check the output mode of the transmitter. See Section 2.3 Wiring the Instrument, Output Select Jumper.
  • Page 31 505-1320 Series Ultrasonic Open Channel Flowmeter 4.6.3 Universal Process Meter (401-90X-2) (continued) 405-1302 Transmitter set in Source mode powers the loop. Figure 4-12 Loop Wiring, Source Mode...
  • Page 32: Section 5 - Troubleshooting

    505-1320 Series Ultrasonic Open Channel Flowmeter SECTION 5 The 505-1302 instrument is designed to give years of unat- TROUBLESHOOTING tended service. No periodic or scheduled maintenance is required. If a problem should occur with the operation of the transmit- Troubleshooting ter, use the following procedure for troubleshooting.
  • Page 33 505-1320 Series Ultrasonic Open Channel Flowmeter Checking the —Operation Check A more precise way to check the temperature sensor is Temperature Sensor to measure the current flow while the unit is on. Refer (continued) to Figure 5-1. µ µ a. Shut power off. b.
  • Page 34 505-1320 Series Ultrasonic Open Channel Flowmeter Table 5-1 Troubleshooting Sheet for Continuous Ultrasonic Instruments Transmitter Model Number __________________Serial Number____________________ Process Material_________________Temperature_______________Pressure_____________ What is the Loop Current? ____________ Is it stable?________________ Test the ability of the electronic unit to produce 4 and 20 mA. Place switch 3 of SW8 to the ON position (up).
  • Page 35: Section 6 Customer Assistance

    505-1320 Series Ultrasonic Open Channel Flowmeter SECTION 6 CUSTOMER ASSISTANCE Factory Assistance AMETEK Drexelbrook can answer any questions about your level measurement system. Call Customer Service at 1-800-553-9092 (US and Canada) , or + 215-674-1234 (International). If you require assistance and attempts to locate the problem have failed: •...
  • Page 36: Equipment Return

    Please include a purchase order even if the repair is under warranty. If repair is covered under warranty, you will not be charged. Ship equipment freight prepaid to: AMETEK-DREXELBROOK. 205 KEITH VALLEY ROAD HORSHAM, PA 19044-1499 COD shipments will not be accepted.
  • Page 37: Section 7 - Specifications

    505-1320 Series Ultrasonic Open Channel Flowmeter SECTION 7 SPECIFICATIONS Transmitter Near/Dead Zone: 12 inches Specifications Minimum Span: 3 inches or 10% of maximum span (whichever is greater) Maximum Span: 87.9 inches Maximum Zero: 99.9 inches Repeatability: 0.1 inch Resolution: 0.1 inch Linearity: 0.5% of full scale for spans above 3 feet.
  • Page 40 205 Keith Valley Road Horsham, PA 19044 US Sales 1-800-553-9092 24 Hour Service 1-800-527-6297 International + 215-674-1234 DREXELBROOK + 215-674-2731 E-mail drexelbrook.info@ametek.com An ISO 9001 Certified Company www.drexelbrook.com...