Working From the Schedule Tab

Use the Schedule tab to schedule routine Zero.
Figure 3-11.
Schedule tab.
Abort Button
Zero Now button
Adjust Offset check box
Set Clock (Time Synchronization)
Zero Duration
3-20 | 5920 UHP Moisture Analyzer
Click to terminate the Zero cycle.
Click to begin Zero cycle now.
Checked
The analyzer performs an offset adjustment at the end of the Zero
cycle.
Not checked
The analyzer verifies only.
The 5920 will always adjust the offset after a Zero cycle.
Click Set Clock to synchronize the clock within the analyzer with the PC.
The Time Synchronization box opens with the PC time and date and the
analyzer time and date. Click Synchronize to set the time, or click Cancel
to close the box.
Enter the Zero duration in cycles. One cycle is equal to 2.5 minutes. The
system defaults to the minimum time required.