6. Feed the strain relief cap with compression bushing onto the new cord and
7. Reconnect 1 (L) and 2 (N) wires (Figure 2-6) by screwing in the two (2) plug
8. Connect a proper, locally recognized plug or terminator to the end of your
9. Replace rear cover.
10. Reconnect power.
Attach ground wire here.
Figure 2-6.
Attach ground wire to
threaded chassis stud.
PN 592047901 Rev YC
tighten securely as per original. If the new cord is a smaller diameter than
the original, consider using the alternate bushing that is included with the
European cord. Ensure that the power cord does not slip when pulled or
pushed from the outside.
screws. Make the ground connection to the threaded chassis stud (Figure
2-7). Do not use the ground position on the power connector. Plug in the
power connector.
cord.