  • Page 1 5100P Moisture Analyzer USER MANUAL...
  • Page 2 This manual is a guide for the use and servicing of the Model 5100P Analyzer. Data herein have been verified and validated and are believed adequate for the intended use of this instrument. If the instrument or procedures are used for purposes over and above the capabilities specified herein, confirmation of their validity and suitability should be obtained otherwise, AMETEK does not guarantee results and assumes no obligation or liability.
  • Page 3 Start Up ........................20 4.2.2 Navigating through the Menu Map ................ 21 USB Interface ........................25 4.3.1 Installing the USB Software ..................25 4.3.2 Connect the 5100P to a Computer Via the USB Cable ........... 28 4.3.3 Navigating the USB Interface .................. 29...
  • Page 4 Environmental Information (WEEE) ..................42 Appendix B - Drawings ........................43 Appendix C - Replacement/Spare Parts ..................47 Spare Parts for 5100P Analyzer ....................47 Spare Parts for Sample Panel ....................47 Contacting AMETEK Service ...................... 48 Appendix D – Abbreviations & Glossary ..................49 Abbreviations ..........................
  • Page 5 Introduction How to Use this Manual Read this manual before operating the 5100P Moisture Analyzer. Failure to do so, or using it in a manner not specified in this manual or any supplemental documents, may reduce the protection against fire, electrical shock, and injury originally provided by this equipment. In addition, failure to follow the installation and start-up instructions may void the instrument’s...
  • Page 6 A note contains important information used to remind or expand on NOTE procedures or processes. It could also contain an action hint or tip. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 7 CONNECTEZ USB OU ALIMENTATION ELECTRIQUE, S’ASSURER QUE L’EMPLACEMENT EST DESIGNE NON DANGEREUX. • If the vent on the rear of the instrument becomes damaged, it must be replaced before further use of the instrument. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 8 In general, for optimum protection against high frequency transients and other disturbances, do not allow installation of the 5100P Analyzer near devices known to generate harmful electromagnetic disturbances (for example, motors, capacitor bank switching, storm-related transients, RF welding equipment, static, walkie-talkies, etc.).
  • Page 9 1.2.4 Warning Statements and Labels These symbols may be seen on AMETEK instruments or in this document to alert you to current or potentially hazardous conditions. PROTECTIVE CONDUCTOR TERMINAL CAUTION - Risk of electric shock CAUTION - Refer to the text accompanying the symbol, and/or any...
  • Page 10 This page is intentionally left blank. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 11 Figure 1 - AMETEK 5100P Operating Modes The 5100P can operate either with or without a Sample Panel. The analyzer also offers a direct user interface through the analyzer display, or enhanced functionality via installed application software and your desktop or laptop computer (USB connection). These are covered on the following pages: •...
  • Page 12 (for example, dryers, replacement sensors, etc.), rapid response, fast recovery from an over-range condition, and long life. Within the 5100P, a TDL is used to produce energy at a very specific wavelength which water molecules strongly absorb. The radiant energy produced by the TDL passes through the sample gas, and any water molecules present in the gas stream will absorb a fraction of this energy.
  • Page 13 Concentration Measurements Once the 5100P has been connected to a sample gas supply and turned on, it cycles through a boot-up sequence. When that is complete, the analyzer starts making measurements on the sample gas. The default setting for the 5100P is to display the moisture concentration measurements in units of parts per million by volume (ppmv), with the sample gas dew point (in degrees Celsius) displayed on the second line.
  • Page 14 This page is intentionally left blank. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 15 5100P Basic User Interface information in Operation of the 5100P Analyzer, page 20. The 5100P is supplied with a Sample Panel, attached to the top of the analyzer, which provides pressure reduction, filtering, and flow control adjustment. A sample gas supply line is connected to the panel, using an 1/8-inch Swagelok®...
  • Page 16 Figure 4 - Analyzer in Vertical Position CAUTION! For Operation in a Hazardous Area: 1. The 5100P must NOT be connected to either a computer via the USB or the battery charger. 2. Protective caps for the charger and USB connectors must be in place for the analyzer to meet the HAZLOC and ingress protection ratings.
  • Page 17 8. Adjust the Flowmeter Valve to set the sample gas flow to between 1 and 2 LPM (see Figure 3 for Flowmeter with Valve location). 9. Once the flow rate of the sample gas is stabilized to 1-2 LPM through the 5100P, the analyzer is ready to measure moisture concentration in the gas stream. See Basic Analyzer Operating instructions on page 20.
  • Page 18 AMETEK recommends a 2-micron particulate filter and a membrane filter capable of removing any liquids, including droplets. Failure to provide a properly conditioned sample gas can result in physical damage to the 5100P and may also result in incorrect moisture concentration measurements.
  • Page 19 • using USB Application software 5100P Power Source Although the 5100P is designed for battery operation, it is important to ensure that the battery pack has been fully charged prior to using the analyzer. 1. To charge the battery pack, connect the 5100P Analyzer to the charger provided.
  • Page 20 6. The 5100P should remain on the mains power for a minimum of eight hours, prior to using the analyzer from the battery pack. If required, the 5100P can remain connected to the charger for an indefinite period. 4.1.1 Extending Battery Life...
  • Page 21 • The Numeric keys are used to enter values. 4.2.2 Navigating through the Menu Map To move through the 5100P menu, use the menu keys on the front panel. The Enter Escape buttons are used to move right and left through the menu. The Enter key moves into the lower levels of the Menu Map (toward application details at the right side of the map).
  • Page 22 New?<Value> Figure 7 - Menu Map Alarm The 5100P can be configured to enable concentration alarm messages. These messages are visible on the display and are also recorded in the data log. Use the following steps to set alarm limits.
  • Page 23 For Dew/Frost point in Centigrade, enter process pressure values in units of Bar absolute. b. For Dew/Frost point in Fahrenheit, enter process pressure values in units of PSIA. 7. Press Enter to save your changes and return to the Main Screen. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 24 USB application. A total of 1440 data points can be maintained in the log. With the default logging interval of one minute, the 5100P can internally store 24 hours of data. After 24 hours (1440 data points) have been stored in the internal data log, the instrument starts to overwrite the oldest data points.
  • Page 25 1. Locate the USB drive supplied with the 5100P Moisture Analyzer. 2. Insert the USB drive into the USB port on the computer to connect to the 5100P. 3. Navigate to the USB drive/device and click to open.
  • Page 26 8. The Wizard sets a file folder into which it will install the software. If you would prefer a different location, click Browse. Choose a different folder or make and name a new Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 27 9. As the software is installing, the Status line fills and updates are noted above the Status line. Wait for the installation to complete. When the installation is complete, the Finish window appears. Figure 12 - Setup Wizard Finish Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 28 10. Click Finish to close the Setup Wizard. 4.3.2 Connect the 5100P to a Computer Via the USB Cable The USB connector is located on the rear panel of the 5100P. Do not connect the 5100P using the USB cable in Hazardous Areas.
  • Page 29 4.3.3 Navigating the USB Interface When the USB application starts up, the Home screen opens (see Figure 15). The 5100P shows on the Home screen with two temperature values displayed. • the temperature of the cell/sample gas is displayed in the bubble to the left of the instrument •...
  • Page 30 1. Using the tabs to other screens (Home, Alarms, Trends, Spectra, Setup) at the left of the screen, click Trend. A Moisture Concentration graph opens. 2. To download and view historical data from the analyzer, select Trend from the Update Interval drop-down menu. Click Apply. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 31 4. To stop the live trending data, click Stop Logging. CAUTION! Collecting live trend data requires the analyzer be connected to a computer. Only access this functionality when the computer and analyzer are in a non-hazardous and safe or general-purpose area. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 32 2. Using the Upload Interval drop-down menu, set an upload interval. Click Apply to view the spectra data. 3. To save spectra data, click Auto Save On at the bottom right of the screen. 4. To stop saving spectra data, click Auto Save Off. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 33 Configure the 5100P Using Setup The Setup tab offers a window that shows system version information and allows configuring several parameters, alarms, and moisture measurement units. Figure 19 - Setup Tab 1. Click the Setup tab on the left side of the screen. A window opens to show System version information, Parameters fields, Alarms fields, a Background Gas field and Units fields.
  • Page 34 5. Units text fields – Select Concentration Units and Dew Point Units from the drop-down lists. For more information on units, see Product Overview – Dew Point Conversions and Concentration Measurements, page 12. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 35 This will prevent any liquids and particulate matter from entering the instrument. When not connected to sample lines and vents, the inlet and outlet connections of the 5100P and Sample Panel should be capped to prevent particulates from entering the Sample Panel or the analyzer.
  • Page 36 The System Fault alarm is active when the 5100P detects a failure in the detector electronics. Data Invalid The Data Invalid alarm is active when the 5100P is not properly locked on the absorption line. It is possible to see this alarm during startup or if the analyzer is exposed to a sudden change in the ambient temperature.
  • Page 37 Technical & Product Specifications Product Dimensions Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 38 *Note: Use only AMETEK supplied Charger for charging the 5100P battery. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 39 Battery Charger Input - 90-264 V AC; 47-63 Hz; 0.62 Amp Battery Charger Output - 7.5 V DC; 4 Amps Environmental Rating Pollution degree:2 Overvoltage category: I Maximum Altitude: 2000 meters Ingress Rating: IP 65 RoHS Compliant Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 40 (90) days from date of shipment. Resale items warranty is limited to the transferable portion of the original equipment manufacturer’s warranty to AMETEK. If you are returning equipment from outside the United States, a statement should appear on the...
  • Page 41 SEPARATELY OR IN CONJUNCTION WITH EQUIPMENT OF OUR MANUFACTURE. WE DO NOT AUTHORIZE ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR US ANY LIABILITY IN CONNECTION WITH EQUIPMENT, OR ANY PART THEREOF, COVERED BY THIS WARRANTY. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 42 Environmental Information (WEEE) This AMETEK product contains materials that can be reclaimed and recycled. In some cases, the product may contain materials known to be hazardous to the environment or human health. To prevent the release of harmful substances into the environment and to conserve our natural resources, AMETEK recommends that you arrange to recycle this product when it reaches its end of life.
  • Page 43 Appendix B - Drawings Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 44 Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 45 Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 46 This page is intentionally left blank. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)
  • Page 47 Appendix C - Replacement/Spare Parts Spare Parts for 5100P Analyzer Part Description Part Number Fiber Panel Protection Cover 7001-536-KE Sealing Washers for Inlet/Outlet Fittings 31765JE Inlet/Outlet Compression Fittings 33354JE Cell Inlet Tube 7001-646-KE Cell Outlet Tube 7001-647-KE Dust Cap for USB Port...
  • Page 48 Contacting AMETEK Service The AMETEK Process Instrument Service Department for this product is located at the US Headquarters in Pittsburgh, Pennsylvania, USA. Normal hours of operation are 9:00 AM to 5:00 PM Eastern Time, Monday through Friday. The Service Department phone number is: 800-537-6044.
  • Page 49 A rating indicating the degree of protection against leakage of harmful dust or water into the analyzer or Sample Panel. Micron A unit of length measurement Sample Spectra Data Measurement data provided by the TDLAS. Revision C 5100P Moisture Analyzer Manual (9000-246-VE)