Variables Table - Ametek 5920 UHP User Manual

Moisture analyzer for ultra-high purity gases
Table of Contents

Advertisement

Variables Table

Name
20mA Range
4ma Range
Baud
SerialMode
NodeAddress
AnalyzerName
Gas
Alarm1Hi
Alarm1Low
KFactor
SerialNumber
CellSerialNumber
Version
SystemReset
HoldOut
EnableAlarm1
ModelName
DryerDateCode
DryerPpbHours
CurrentPpbHours
LCDcontrast
LCDbacklight
AdjustFreqOffset
ZeroHour
ZeroPeriod
ZeroMonthDays
ZeroWeekDays
ZeroType
CellTemp
AnalogOutput
EETemp
Moisture
HeldConc
Frequency
DeltaFrequency
Flow1
Flow2
AirTemperature
Figure 2-19.1.
Variables table.
2-30 | 5920 UHP Moisture Analyzer
ID
Description
1D
High current output limit
1E
Low current output limit
28
Serial Communication Speed
29
RS-485 two-wire or four-wire mode
2A
RS-485 analyzer address
2B
Ametek Analyzer Name
2C
Sample gas name
2E
High concentration limit
2F
Low concentration limit
38
Gas based coefficient
3E
Analyzer manufacturing number
3F
Sensor manufacturing code
41
Embedded software version
42
Analyzer reboot command
45
Hold current outputs during Zero cycle
49
Enable or disable concentration alarm
77
Analyzer name
4B
Dryer production code
4C
Dryer consumption in 1000 of ppB hours
4D
Dryer consumption in ppB hours
4F
Display contrast setting
50
Display back light on/off setting
51
Adjust frequency offset after Zero
52
Automatic Zero schedule parameter
53
Number of cycles to stay in Zero
54
Automatic Zero schedule parameter
55
Automatic Zero schedule parameter
56
Automatic Zero type(daily, monthly, weekly)
58
Sensor temperature
59
Current outputs reading
5A
Electronics enclosure temperature
5B
Moisture concentration reading
5C
Held moisture concentration reading
5D
Sensor frequency reading
5E
Wet minus Dry reading
5F
Reference channel flow
60
Total flow
62
Oven air temperature
Type
Typical Value
Float
100.0
Float
0.0
Integer
9600
Integer
1 (four-wire)
String[4]
1
String[41] 5920
String[41] Air
Float
100.0
Float
0.0
Float
1.0
String[13] 123
String[13] 591015
String[4]
1.00
Integer
0
Integer
1(hold)
Integer
1(enable)
String[16] AMETEK 5920 UHP
String[10] 010080009
Float
200000.00
Float
124.00
Integer
5
Integer
1(on)
Integer
1(adjust)
Integer
23
Integer
24
Integer
26
Integer
1
Integer
0(daily)
Float
50.0 degreeC
Float
12.0 mA
Float
55.0 degreeC
Float
20.0 ppB
Float
20 ppB
Float
5300 Hz
Float
0.0 Hz
Float
60.0 sccm
Float
120 sccm
Float
60.0 degree C
PN 592047901 Rev YC

Advertisement

Table of Contents
loading

This manual is also suitable for:

5910