Power Installation

If the factory-supplied flexible power cord is not long enough or is not compli-
ant with local standards, a 3-meter European-type Harmonized power cord is
also included.
Ensure that the area is safe and replace the power cord as follows:
1. Disconnect the power from the analyzer and remove the Rear Panel termi-
2. Unplug the power connector (Figure 2-5) and unscrew the two (2) plug
3. Remove the ground wire from the threaded chassis stud (Figure 2-6).
4. Loosen and remove the power cord strain relief cap with compression
5. Prepare the customer supplied power cord:
2 (N) (NEUTRAL)
Figure 2-5.
Remove 1 (L) and 2 (N).
2-12 | 5920 UHP Moisture Analyzer
nal cover.
screws that hold the "1" (LINE) and "2" (NEUTRAL) wires in place.
bushing (Item 1, Figure 2-2) and pull out the power cord.
-
remove outer jacket
-
measure length of wires
-
strip lengths of wires
- ring connector termination, etc.