Performing an Initial Zero

To start the initial Zero cycle:
1. Press the Zero key.
2. Press the '8' key (arrow down) until you see "Zero" between the arrows.
This initial Zero cycle takes approximately two (2) hours. When the Zero cycle
is complete, the analyzer will adjust the offset and automatically return to
normal sampling.
PN 592047901 Rev YC
Select "Zero" by pressing the Enter key.
The 5920 Analyzer requires time to fully recover to normal operation
after a Zero cycle. During the recovery period, "Equilibrating" will be
shown on line two of the display.
Installation and Start-Up
|
2-23