Range Key - Ametek 5920 UHP User Manual

Moisture analyzer for ultra-high purity gases
Table of Contents

Advertisement

Range Key

Use the Range key to define your 4–20 mA output setting.
Use this setting to scale the 4–20 mA output. The 4–20 mA output is propor-
tional to the moisture readings.
Figure 3-5.
Main Menu, Range key.
Maximum Conc.
Minimum Conc.
Example:
To scale the analog output to represent a 0–100 PPMV range, enter the follow-
ing settings:
Maximum Conc.:
Minimum Conc.:
PN 592047901 Rev YC
RANGE
Enter the moisture reading equivalent to 20 mA.
Enter the moisture reading equivalent to 4 mA.
100
0
Maximum Conc
Enter Value
Minimum Conc
Enter Value
Controller/Interface
|
3-5

Advertisement

Table of Contents
loading

This manual is also suitable for:

5910

Table of Contents