5830 Moisture Analyzer Specifications - Ametek 5830 User Manual

Moisture analyzer
Hide thumbs Also See for 5830:
Table of Contents

Advertisement

5830 Moisture Analyzer Specifications

Specification
Range
Limit of Detection
Accuracy
Response Time
Sample Inlet Pressure
Requirement
Exhaust Pressure
Sample Flow Requirement Approximately 150 SCCM in sample saver mode.
Inlet Gas Temperature
Outputs
Alarms/Alerts
Environmental Conditions
Utility Requirements
5-2 | 5830 Moisture Analyzer
Description
Calibrated range of 0–100 PPMV. Trend indication to 1000 PPMV
20 PPBV nominal
± 20 PPBV or ±10% of reading, whichever is greater
Typical 63% of a step change in either direction in less than 5
minutes
High Pressure
138 kPa to 345 kPa ( 20 PSIG to 50 PSIG )
Low Pressure:
34.5 kPa to 138 kPa (5 PSIG to 20 PSIG) See Table below
Helium
5-20 PSIG (34.5-138 kPa)
Hydrogen
5-20 PSIG (34.5-138 kPa)
Nitrogen
5-20 PSIG (34.5-138 kPa)
Oxygen
5-20 PSIG (34.5-138 kPa)
Argon
7-20 PSIG (48.3-138 kPa)
Atmospheric
Approximately 1 SLPM with bypass open at 20 PSIG.
Approximately 250 SCCM with bypass open at 5 PSIG
0 °C to 100 °C (32 °F to 212 °F)
Optimal results are obtained when inlet gas temperature is
maintained at
60 °C (140 °F).
4-line x 20-character LCD display
One (1) 4–20 mA DC Analog Output with FS ± 1% accuracy
RS-232 / RS-485 serial port
Three (3) independent contact closures of 30 VAC or 60 VDC
maximum, 10 VA or 1 A maximum, resistive for System Alarm,
moisture Concentration Alarm, and Data Valid. All are fail-safe by
default.
Alarms are available on RS-485 interface.
5 °C to 50 °C (41 °F to 122 °F)
[Optimal performance in low-level applications is obtained
when ambient temperature is maintained at a constant
temperature (± 2 °C)]
90% RH maximum, noncondensing
IEC Pollution Degree 2
Maximum altitude 2,000 meters (6560 feet)
IEC Installation Category II
Indoor use only, IEC 529; IP40
Power Consumption
100-127 Vac, 47-63 Hz, 185 Watts max OR
220-240 Vac, 47-63 Hz, 185 Watts max
Instrument Air
483 - 690 kPa (70 - 100 PSIG), -40 °C dew point
5-micron filter recommended
PN 583054901, Rev YA