Compatible Gases

Air, argon, carbon dioxide, custom gas, Freon 12, Freon 114, Freon 116, helium,
hydrogen, krypton, neon, nitrogen, oxygen, SF6, and xenon. Carbon dioxide
requires a custom Sensor assembly. Contact AMETEK Service to confirm com-
patibility with other gases.

3050-AM Moisture Analyzer Specifications

Specification
Range
Limit of Detection
Accuracy
Response Time
Sample Inlet Pressure
Requirement
Exhaust Pressure
Sample Flow Requirement Approximately 50 SCCM
Inlet Gas Temperature
Outputs
PN 305804902, Rev A

Specifications

All specifications subject to change without notice.
Read the Dryer Warning in the "Special Warnings and Information"
section in the beginning of this manual for more information.
Description
Calibrated range of 0–100 PPMV. Trend indication to 1000 PPMV
35 PPMV nominal
±10 % of reading OR 35 PPMV, whichever is greater.
Typical 63 % of a step change in either direction in less than
5 minutes
42 kPa to 345 kPa (6 PSIG to 50 PSIG)
Atmospheric
Approximately 900 SCCM @ 50 psig inlet with Bypass Option
0 °C to 100 °C (32 °F to 212 °F)
Optimal results are obtained when inlet gas temperature is
maintained at 60 °C (140 °F).
4-line x 20-character LCD display
One (1) 4–20 mA DC analog output
RS-232 / RS-485 serial port
|