Solenoid Valves

Two (2) solenoid valves are attached to a solenoid bracket located on the top of
the MFM mounting bracket. These valves are labeled SAM (Figures 4-1, 4-2, and
4-5) for the sample solenoid valve assembly (AMETEK Part No. 591016901) and
REF for the reference solenoid valve assembly (AMETEK Part No. 591092901).
They are connected by wire to the board at the location marked on the side of
the valve wire connector. A third reference solenoid valve (AMETEK Part No.
592094901) is located on the other side of the mounting plate and is attached
to the Reference Purifier. It is mounted on the reference solenoid bracket and
labeled SAM 1.
Figure 4-4.
Sample solenoid valve
assembly.
To remove the Solenoid Valve assembly:
Follow these instructions to remove either the sample or reference solenoid
valve assembly from the 5920 Analyzer.
1. After power has been disconnected, remove the top cover and sample
2. Locate the solenoid valve wires for the valve assembly that you are replac-
4-8 | 5920 UHP Moisture Analyzer
system oven cover from the analyzer and locate the solenoid valves. They
are attached to the solenoid bracket on top of the MFM mounting bracket
(Figure 4-5).
ing and follow the wires to the connector end on the board. Note the loca-
tion where the connector is attached to the board. Remove the connector
from the MCU board.
PN 592047901 Rev YC