Special Addresses
Addresses $F0 through $FF are reserved for special functions with address
$F0 used as a broadcast address.
Address $F0
Node address $F0 is used as a single point address (i.e., it is intended for use in
configurations with a single slave). The slave will respond to all messages with
node address $F0. You must be careful to prevent multiple slaves from receiv-
ing commands with the broadcast address as this will cause a bus contention.
Defined Commands (Master to Slave)
The commands that can be issued from the master are listed in Figure 2-18.
Figure 2-18.
Defined Commands table.
Echo (A)
Echo implements a loop back function to test the integrity of the serial com-
munication. Any text transmitted by the master is returned.
Example:
Master:
Slave:
2-26 | 5920 UHP Moisture Analyzer
Command Byte
Function
A
Echo
B
Bad Command
C
Acknowledge
F
Read Data
G
Not Used
H
Write Data
L
Write Cell EEPROM
Q
Quit Zero
R
Analyzer Reset
T
Toggle Test Mode
Z
Initiate Zero
>99ABCDEFG4E[CR]
AABCDEFG1D[CR]
PN 592047901 Rev YC