Zero Schedule

Use this function to schedule an automatic analyzer Zero cycle.
Figure 3-6.6.
Zero Schedule menu.
Schedule Type
Day of the Week
Day of the Month
Hour
Duration
3-10 | 5920 UHP Moisture Analyzer
Config
Zero Schedule
Choose the preferred Schedule Type:
Never, Daily, Weekly, or Monthly.
Select the day of the week from the list.
Enter the 2-digit value for the day of the month.
Enter the 2-digit value for the hour of the day using a 24-hour clock. Zero is
the starting hour.
Enter the Zero period value in minutes.
AMETEK recommends performing a Zero every 24 hours for the first
month of operation. After this start-up period, the analyzer will have
dried down sufficiently to increase the Zero interval to 1 week. AMETEK
recommends that the duration of the Zero cycle be set to 48 (i.e., 2
hours).
Never
Daily
Weekly
Monthly
Schedule
Type
Sunday
Monday
Tuesday
Day of
Wednesday
Week
Thursday
Friday
Saturday
Day of
Enter Value
Month
Enter Value
Hour
Enter Value
Duration