From these data, the slight loss of sensitivity experienced by the population
of 5900 analyzers is estimated to reach 10 % (i.e., a 2 PPBV change at 20 PPBV)
after 45 months of service.
Even though the QCM-based moisture analyzers are extremely stable, AMETEK
recommends that you check the calibration of the analyzer every three (3)
years. This performance check should be handled as part of the normal main-
tenance schedule for the equipment.