5.2 Basic Settings (
● If preset is turned ON when the preset value is "0", zero-point adjustment will take place.
● A preset value setting change will be applied even if the preset has been executed.
● For details, refer to
Setting method
Setting item
Preset value (PR. VAL)

5.2.7 Preset Data Selection (

When using a preset, you can select the target data (NORM. V or JUDGE.V).
You can select "Judgment value (JUDGE.V)" to use the held value as the preset.

Preset data selection

Judgment value (JUDGE.V)
Normal measured value
(NORM.V)
● When a calculation function is used ("standard difference" on the master unit, or a mode other
than "standard difference" on the slave unit), the setting cannot be changed.
● If a calculation function ("standard difference" on the master unit, or a mode other than
"standard difference" on the slave unit) is set after preset data selection is set to "Judgment
value (JUDGE.V)", the preset data will be the normal measured value (NORM.V).
5-18
)
"4.2.2
Preset".
-199.9999 to 199.9999 (mm)
)
The judgment value (JUDGE.V) is used as the preset value.
The normal measured value (NORM.V) is used as the preset value.
Setting range
Function
Default value
0.0000
WUME-HGS-11