5.7 Calibration Settings (
5.7 Calibration Settings (
Structure of calibration setting menu
5.7.1 Calibration Selection (
You can perform zero-point adjustment and span adjustment when a sensor head is replaced to
reduce installation error.
Calibration settings
Calibration selection
(CAL.SEL)
5-90
)
)
)
Set values
Default (DEFAUL)
User settings (USER)
Acquired value 1 (CL.SET1)
Acquired value 2 (CL.SET2)
Adjustment value 2 (AJ.VAL2)
Function
Set to the factory default state.
Execute calibration using the preset values
below, and correct the zero point and span.
Move the spindle to the zero point for calibration
and acquire the measured value.
Move the spindle to the target point for span
adjustment and acquire the measured value.
Enter a number for the target value.
WUME-HGS-11