Panasonic HG-S Series User Manual page 104

Contact-type digital displacement sensor
Hide thumbs Also See for HG-S Series:
Table of Contents

Advertisement

5.3 Advanced Settings (
Timing chart
1. The judgment value (JUDGE.V) is held while the external trigger input is ON.
2. The minimum normal measured value (NORM.V) is updated and bottom hold is performed
with this value.
3. If the reset input turns ON while the external trigger input is OFF, an indeterminate state
occurs during the response time. All outputs remain OFF during an indeterminate state.
4. After the response time elapses, the indeterminate state is released.
5. If the reset input turns ON while the external trigger input is ON, the indeterminate state is
held until the external trigger input turns OFF.
6. Judgment output outputs the judgment result based on the output setting.
* When the calculation function is used, calculation value "CALC" is shown instead of normal
measured value (NORM.V).
Peak to Peak Hold (Trigger mode: Hold)
With respect to normal measured values (NORM.V) during any sampling interval, the difference
between the peak value (maximum value) and the bottom value (minimum value) is held as the
judgment value (JUDGE.V).
5-32
)
WUME-HGS-11

Advertisement

Table of Contents
loading