Pressure check set value
Setting item
Pressure check set value (PRS.SET)
● If the setting is changed from the default value when the connected sensor head type is
changed, the original set value will be held.
● If the setting is not changed from the default value when the connected sensor head type is
changed, the default value will be reset according to the sensor head type.
● On an HG-SC series controller manufactured in January 2019 or earlier, the setting is not
reset by simply connecting the sensor head. Always turn the power off and then on after
connecting..
Catch check
If the movement of the spindle worsens due to aging and the spindle does not return
completely, accurate measurement will no longer be possible. This function detects failure of
the spindle to return completely.
WUME-HGS-11
5.3 Advanced Settings (
Setting range
-199.9999 to 199.9999 (mm)
)
Default value
10 mm type:
10.0000
32 mm type:
32.0000
50 mm type:
50.0000
Air-driven type:
10.0000
5-69