R&S
FSV-K8
Parameters:
<Mode>
Example:
Usage:
Mode:
Manual operation:
CALCulate<n>:BTOoth:RTPower:GFSK? <Mode>
This command reads the transmission power for the GFSK sections of the packet in
the Relative Transmission Power measurement. With a sweep count 1, the command
returns the current value, irrespective of the parameter.
Note:
This command is only available with active Relative Transmit Power measurement
(see
ments this command causes a query error.
Suffix:
<n>
Parameters:
<Mode>
Example:
Usage:
Operating Manual 1176.7610.02 ─ 02
MINimum | MAXimum | AVERage
*RST:
For the first steps refer to
ple", on page 127
CONF:BTO:MEAS RTP
Activates the Relative Transmit Power measurement.
CONF:BTO:SWE:COUN 10
Initiates the sweep counter with 10.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:RTP? MIN
Reads the lowest recorded average power of the DPSK section
of all packets.
Query only
BT
See
"Rel TX Power"
CONFigure: B TOoth: M EASurement
.
irrelevant
MINimum | MAXimum | AVERage
*RST:
For the first steps refer to
ple", on page 127
CONF:BTO:MEAS RTP
Activates the Relative Transmit Power measurement.
CONF:BTO:SWE:COUN 10
Initiates the sweep counter with 10.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:RTP:GFSK? MIN
Reads the highest recorded average power of the GFSK section
of all packets.
Query only
Remote Commands for Bluetooth Measurements (K8)
-
chapter 4.14, "Programming Exam-
on page 44
on page 96). With all other measure-
-
chapter 4.14, "Programming Exam-
CALCulate subsystem
86