Format Subsystem; Initiate Subsystem - R&S FSV-K8 Operating Manual

Firmware option bluetooth measurements
Table of Contents

Advertisement

®
R&S
FSV-K8
Manual operation:

4.4 FORMat subsystem

FORMat[:DATA] <Format>
This command selects the data format for the data transmitted from the R&S FSV to
the controlling computer. It is used for the transmission of trace data. The data format
of trace data received by the instrument is automatically recognized, regardless of the
format which is programmed.
Parameters:
<Format>
Example:

4.5 INITiate subsystem

INITiate<n>:CONTinuous
INITiate<n>[:IMMediate]
INITiate<n>:CONTinuous <State>
This command determines whether the trigger system is continuously initiated (contin-
uous) or performs single measurements (single).
The sweep is started immediately.
Suffix:
<n>
Parameters:
<State>
Operating Manual 1176.7610.02 ─ 02
Remote Commands for Bluetooth Measurements (K8)
See
"Range Log Manual"
See
"Range Log 100 dB"
See
"Range Log 50 dB"
See
"Range Log 10 dB"
See
"Range Log 5 dB"
See
"Range Log 1 dB"
See
"Range Linear %"
See
"Range Lin. Unit"
ASCii
ASCII data are transmitted in plain text, separated by commas.
REAL
REAL data are transmitted as 32-bit IEEE 754 floating-point
numbers in the "definite length block format".
*RST:
ASCII
FORM REAL,32
FORM ASC
................................................................................................109
..................................................................................................110
.
irrelevant
ON | OFF
*RST:
ON
FORMat subsystem
on page 49
on page 50
on page 50
on page 50
on page 50
on page 51
on page 51
on page 51
109