[SENSe:]SWEep:MODE <Mode>
This command selects the spurious emission and spectrum emission mask measure-
ments.
Parameters:
<Mode>
Example:
[SENSe:]SWEep:POINts <NumberPoints>
This command defines the number of measurement points to be collected during one
sweep.
Parameters:
<NumberPoints>
Example:
[SENSe:]SWEep:TIME <Time>
This command defines the sweep time.
The range depends on the frequency span.
Parameters:
<Time>
Example:
[SENSe:]SWEep:TIME:AUTO <State>
In realtime mode, this command automatically sets the sweep time to 32 ms.
In analyzer mode, this command controls the automatic coupling of the sweep time to
the frequency span and bandwidth settings. If
matic coupling is switched off.
Operating Manual 1176.7610.02 ─ 02
Remote Commands for Bluetooth Measurements (K8)
AUTO | ESPectrum | LIST
AUTO
Switches to "Spectrum" measurement mode or stays in the cur-
rent mode if it is not ESP/LIST
ESPectrum
"Spectrum Emission Mask" measurement mode
LIST
"Spurious Emissions" measurement mode
*RST:
AUTO
SWE:MODE ESP
Sets the Spectrum Emission Mask measurement mode.
Range:
101 to 32001
*RST:
691
SWE:POIN 251
refer to data sheet
*RST:
(automatic)
SWE:TIME 10s
SENSe Subsystem (BLUETOOTH, K8)
[SENSe:​ ] SWEep:​ T IME​
is used, auto-
118