Configure:btooth:sweep:count - R&S FSV-K8 Operating Manual

Firmware option bluetooth measurements
Table of Contents

Advertisement

®
R&S
FSV-K8
CONFigure:BTOoth:SWEep:COUNt <NoSweeps>
This command defines the number of sweeps for the currently selected Bluetooth mea-
surement. This number will be started as a "single sweep". With trace mode setting
clear/write (see
ment results will be calculated for each sweep and taken into account for the calcula-
tion of minimum, maximum and average values. For the remaining trace settings
(AVER, MAXH, MINH) the measurement results will be calculated from the resulting
trace. In average mode the value 0 defines the sliding average of the measurement
data over 10 sweeps.
Note:
The setting is valid only for the currently active measurement (see
BTOoth:​ M EASurement​
ments.
Parameters:
<NoSweeps>
Example:
Mode:
Manual operation:
Table 4-3: RST value depending on measurement type
Measurement type
Output Power
Adjacent Channel Power
Modulation Characteristics
Initial Carrier Frequency Tolerance
Carrier Frequency Drift
EDR Relative TX Power
EDR In-band Spurious Emissions
Differential Phase Encoding
Operating Manual 1176.7610.02 ─ 02
Remote Commands for Bluetooth Measurements (K8)
CONFigure:BTOoth Subsystem (BLUETOOTH, K8)
CONFigure:​ B TOoth:​ T RACe<t>:​ M ODE​
on page 96) and independent of the other Bluetooth measure-
0 to 32767
*RST:
see table below
INST:SEL BTO
Activates the Bluetooth Measurements option.
CONF:BTO:MEAS OPOW
Activates the Output Power measurement.
INIT:CONT OFF
Selects single sweep operation.
CONF:BTO:SWE:COUN 20
Sets the number of sweeps to 20.
INIT;*OPC
Starts a sweep with synchronization.
BT
See
"Sweep Count"
RST value
0
10
100
on page 103) the measure-
on page 57
CONFigure:​
102

Advertisement

Table of Contents
loading

Table of Contents