Spectrum Analyzer ¸FSH
¸FSH3 100 kHz to 3 GHz
¸FSH6 100 kHz to 6 GHz

Advertisement

loading

Summary of Contents for R&S FSH3

  • Page 1 Handheld Spectrum Analyzer ¸FSH ¸FSH3 100 kHz to 3 GHz ¸FSH6 100 kHz to 6 GHz...
  • Page 2 Spectrum analysis anywhere, anytime – on earth and in space The ¸FSH is the ideal spectrum analyzer for rapid, high-precision, cost-effective signal investi- gations. It provides a large number of measure- ment functions and so can handle anything from the installation or maintenance of a mobile radio base station up to on-site fault location in RF cables as well as development and service –...
  • Page 3 Handy, robust and portable Trace  Memory Trace  Clear/Write  Max/Min Hold  Average  View  Detectors  – Auto Peak – Sample – Max/Min Peak – RMS The ¸FSH has been designed as a robust, portable spectrum analyzer that can be used in Function keys Softkey function the field.
  • Page 4 Selection of measurement functions: Spectrum analysis  Scalar network analysis  Vector network analysis  Receiver mode  Channel power  TDMA power  Occupied bandwidth   3GPP code domain power  Isotropic antenna  Selection of following functions: C/N measurement ...
  • Page 5 AC power supply connector Generator output, N connector Power sensor connector Trigger input/ external reference input RF input, N connector Data in brief ¸FSH 3 ¸FSH 6 Headphones connector Frequency range 100 kHz to 3 GHz 100 kHz to 6 GHz Resolution bandwidths 1 kHz to 1 MHz (model .13)
  • Page 6 Not available ¸FSH – models Frequency range Tracking generator Output power of Preamplifier Resolution bandwidth tracking generator ¸FSH3 model .03 100 kHz to 3 GHz – – 100 Hz to 1 MHz  ¸FSH3 model .13 100 kHz to 3 GHz –20 dBm...
  • Page 7 TDMA power measurements By means of the TDMA POWER function, the ¸FSH performs time-domain power measurements within a timeslot of TDMA (time division multiple access) methods. All the settings required for the GSM and EDGE standards are predefined on the ¸FSH to make these measurements easier for the user. In addition, up to five user-definable instrument setups can be loaded into the ¸FSH using the ¸FSH View software.
  • Page 8 C/N measurements The ¸FSH offers a carrier/noise (C/N) measurement for determining the ratio of carrier power to noise power or carrier power to noise power density. The ¸FSH supports three different modes for carrier power measurement. In the CW TX mode, the ¸FSH determines the power of an unmodulated carrier.
  • Page 9 Power measurements The Power Sensors ¸FSH-Z1 and ¸FSH-Z18 expand the ¸FSH to a high-precision RF power meter up to 8 GHz and 18 GHz respectively. As with thermal sensors, the true RMS value of the measured signal is obtained over the entire measurement range of –67 dBm to +23 dBm irrespective of the signal waveform.
  • Page 10 Scalar transmission and reflection measure- ments with VSWR bridge (¸FSH-Z2/-Z3 as accessory) The ¸FSH with built-in tracking generator rapidly determines the transmission characteristics of cables, filters, amplifiers, etc, with a minimum of effort. When equipped with the ¸FSH-Z2/-Z3 VSWR bridge (10 MHz to 3 GHz/6 GHz), the ¸FSH can also measure the matching (return loss, reflection coefficient or VSWR) of an antenna, for example.
  • Page 11 3GPP FDD code domain power measurements on base stations The ¸FSH-K4 option allows the code domain power measurements on a 3GPP base station. It measures the total power and the power of the most important code channels, such as the common pilot channel (CPICH), primary common control physical channel (P-CCPCH), primary synchronization channel (P-SCH) and secondary synchronization channel (S-SCH).
  • Page 12 Control Software ¸FSH View The powerful software package for documenting your measurements is supplied with every ¸FSH. Features Runs under Windows 98/ME/NT/2000/XP Automatic storage of measurement results at selectable   intervals Rapid and simple transfer of measurement data from the ...
  • Page 13: Specifications

    Specifications apply under the following conditions: 15 minutes warm-up time at ambient temperature, speci- fied environmental conditions met and calibration cycle adhered to. Data without tolerances: typical values. Data designated as “nominal”: design parameters, i. e. not tested. ¸FSH3 ¸FSH6 Frequency...
  • Page 14 ¸FSH3 ¸FSH6 Amplitude Display range average noise level displayed to +20 dBm Maximum permissible DC voltage at RF 50 V/80 V input Maximum power 20 dBm, 30 dBm (1 W) for max. 3 minutes Intermodulation-free dynamic range third-order IM products, typ.
  • Page 15 ¸FSH3 ¸FSH6 Markers Number of markers or delta markers max. 6 Marker functions peak, next peak, minimum, center = marker frequency, reference level = marker level, all markers to peak Marker displays normal (level), noise marker, frequency counter (count) Trigger...
  • Page 16 ¸FSH3 ¸FSH6 Accessories Accessories Power Sensors ¸FSH-Z1 and ¸FSH-Z1 Frequency range ¸FSH-Z1 10 MHz to 8 GHz ¸FSH-Z18 10 MHz to 18 GHz VSWR 10 MHz to 30 MHz <1.15 30 MHz to 2.4 GHz <1.13 2.4 GHz to 8 GHz <1.20...
  • Page 17 ¸FSH3 ¸FSH6 Temperature coefficient 25 MHz to 40 MHz 0.40 %/K (0.017 dB/K) 40 MHz to 1 GHz 0.25 %/K (0.011 dB/K) Peak envelope power Power measurement range for video bandwidth 4 kHz 0.4 W to 300 W 200 kHz...
  • Page 18 ¸FSH3 ¸FSH6 Directional Power Sensor ¸FSH-Z Frequency range 200 MHz to 4 GHz Power measurement range 30 mW to 120 W (300 W with unmodulated envelope) VSWR referenced to 50 W <1.07 200 MHz to 3 GHz <1.12 3 GHz to 4 GHz...
  • Page 19 Load matching Return loss 200 MHz to 3 GHz 0 dB to 23 dB 3 GHz to 4 GHz 0 dB to 20 dB VSWR 200 MHz to 3 GHz >1.15 3 GHz to 4 GHz >1.22 PEAK (max. 10 ms), VSWR ≤ 3 1000 AVG –10°C to +35°C VSWR ≤...
  • Page 20 169 mm × 116 mm × 30 mm 149 mm × 144 mm × 45 mm Weight 485 g 620 g Distance-to-Fault Measurement option ¸FSH-B1 only with ¸FSH3 models .13/.23/.26 and ¸FSH-Z2/-Z3 VSWR bridges Display 301 pixel Maximum resolution, distance to fault maximum zoom...
  • Page 21 ¸FSH3 ¸FSH6 Transmission measurements (only with ¸FSH3 models .13, .23 and ¸FSH6 model .26) Frequency range 5 MHz to 3 GHz 5 MHz to 6 GHz Dynamic range 10 MHz to 2.2 GHz scalar mode typ. 60 dB typ. 80 dB vector mode, option ¸FSH-K2...
  • Page 22 ¸FSH 3 ¸FSH 6 3GPP FDD code domain power BTS/Node B measurement (only with R&S®FSH-K 1300.633.02 and ¸FSH 3 model .23) Frequency range 10 MHz to 3 GHz – Carrier frequency uncertainty (test case 6.3 in line with – 3GPP 25.141) Measurement range ±1 kHz –...
  • Page 23: General Data

    General data Display 14 cm (5.7“) LC color display Resolution 320 × 240 pixel Memory CMOS RAM Settings and traces up to 256 Environmental conditions Temperature Operating temperature range ¸FSH powered from internal battery 0 °C to +50 °C ¸FSH powered from AC power supply 0 °C to +40 °C Storage temperature range –20 °C to +60 °C...
  • Page 24: Ordering Information

    ¸FSH-K1 1157.3458.02 Vector Transmission and Reflection Measurements ¸FSH-K2 1157.3387.02 Receiver Mode ¸FSH-K3 1157.3429.02 3GPP FDD Code Domain Power BTS/Node B Measurement for R&S®FSH3 model .23 ¸FSH-K4 1300.7633.02 Power sensor connector on ¸FSH3 Power Sensor ¸FSH-Z1 VSWR Bridge and Power Divider ¸FSH-Z2 For ¸FSH 3 only (1145.5850.23), as of serial no 103500.
  • Page 25: Optional Accessories

    Spare CD-ROM with Control Software ¸FSH View and documentation ¸FSH-Z35 1145.5821.02 Power sensor Spare Headphones ¸FSH-Z36 1145.5838.02 connector on ¸FSH3 Spare USB Optical Cable ¸FSH-Z37 1300.7733.02 Matching Pad 50/75 W, 0 Hz to 1000 MHz ¸FSH-Z38 1300.7740.02 Active Directional Antenna ¸HE200...
  • Page 26 Certified Quality System Certified Environmental System ISO 9001 ISO 14001 DQS REG. NO 1954 QM DQS REG. NO 1954 UM More information at www.fsh.rohde-schwarz.com Rohde &Schwarz direct: Tel. (+49 2203) 807-800 Fax (+49 2203) 807-66 E-mail: Direct@rohde-schwarz.com www.rohde-schwarz.com Europe: +49 1805 12 4242, customersupport@rohde-schwarz.com USA and Canada: +1-888-837-8772, customer.support@rsa.rohde-schwarz.com Asia: +65 65 130 488, customersupport.asia@rohde-schwarz.com...