CONFigure:BTOoth:PTYPe <Type>
This command selects the packet type to be measured.
Parameters:
<Type>
Example:
Mode:
Manual operation:
CONFigure:BTOoth:RTPower:DAVerage:STARt <Time>
This command sets the start time for the power measurement of the DPSK sections of
the packet.
Note:
This command is only available with active Relative Transmit Power measurement
(see
Parameters:
<Time>
Example:
Mode:
Manual operation:
CONFigure:BTOoth:RTPower:DAVerage:STOP <Time>
This command sets the stop time for the power measurement of the DPSK sections of
the packet.
Note:
This command is only available with active Relative Transmit Power measurement
(see
Operating Manual 1176.7610.02 ─ 02
DH1
1 slot packet
DH3
3 slot packet
DH5
5 slot packet
AUTO
automatic selection of the packet type
*RST:
CONF:BTO:PTYP DH5
Selects type "5 slot packet".
BT
See
"Packet Type"
CONFigure: B TOoth: M EASurement
0 to 100%
*RST:
CONF:BTO:RTP:DAV:STAR 20
Sets the start time for the power measurement of the DPSK sec-
tions of the packet.
BT
See
"DPSK Start"
CONFigure: B TOoth: M EASurement
Remote Commands for Bluetooth Measurements (K8)
CONFigure:BTOoth Subsystem (BLUETOOTH, K8)
DH1
on page 36
on page 96).
10%
on page 45
on page 96).
100