Verifying DigiSizer System Components
Verifying DigiSizer System Components
Procedure
Enter service test mode.
Leak check the DigiSizer.
Move rotation arm to the specified angle.
Fire laser.
Ensure that temperature is within band.
Record the application version of the sample handler:_______________
Check temperature of fluid.
Test analysis liquid pump. Add warm water and ensure that the
temperature reading on the screen changed.
Test rinse pump.
Test circulating pump.
Test waste valve.
Test ultrasonic probe.
Leak check the AquaPrep.
Test the AquaPrep flow rate. Elapsed time should be within 20 to 60
seconds.
92
Saturn DigiSizer II 5205
Evaluation Code
Initial
(circle one)
/Date
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A
P N F N/A