Micromeritics Saturn II 5205 DigiSizer Installation Instructions And Checklist page 74

Table of Contents

Advertisement

Verifying Operation
8. Select Show instrument schematic, then Enable manual control from the Unit menu.
Circulating Pump
9. Right-click on the circulating pump; choose Set speed. Enter 1500 if you have the
standard liquid sampler installed, or 4000 if you have the Low-volume liquid sampler
installed, then click
10. Right-click again and turn on the circulating pump while observing the sample cell.
• If leaks occur, turn off the pump and open the drain valve immediately. Dry the sample
cell area, correct the problem, and repeat the steps again.
• If leaks do not occur after approximately 10-15 seconds, turn off the pump and open the
drain valve to drain the liquid from the system.
11. Turn the pump back on for 3 to 5 seconds to expel any additional liquid that may have
remained in the pump, then turn it off.
12. Close the drain valve.
68
OK
.
Saturn DigiSizer II 5205
MasterTech portion of
schematic
(shown only if a
MasterTech is installed)
Jul 09