Appendix A
Moving an Existing Configured Analyzer to Another Computer
Use the instructions in this section if you have a configured analyzer that you wish to move to
a different computer that already has an analyzer and the analysis program installed.
1.
Review the Saturn DigiSizer Configuration forms for both the current and destination
computers. If there are duplicate IP addresses for analyzers (in most cases, there will be),
it is best to change the IP address using the current computer before moving the analyzer.
a. Turn on installed analyzers and start the analysis program.
b. Select Unit configuration from the Unit [n] menu and click
Setup dialog is displayed.
c. Enter a new address for the analyzer you are uninstalling and record it on the Saturn
DigiSizer Configuration form for the destination computer. For example, the current
analyzer is nnn.nnn.nnn.101 and you have an analyzer already installed on the
destination computer with the same IP address. Specify the address for the analyzer
you are uninstalling as nnn.nnn.nnn.102.
d. Click
dialog.
2.
Exit the analysis program and turn off all analyzers.
3.
Disconnect the Ethernet cable of the analyzer from the current computer (or Ethernet
switch) and reconnect to the Ethernet switch for the destination computer.]
4.
On the current (source) computer, select Start from the Task bar, and Run from the Start
menu.
5.
Enter the drive designator of the CDROM drive, followed by SETUP; for example,
e:setup.
6.
Click
OK
A-6
to close the Unit IP Setup dialog, then again to close the Unit Configuration
OK
; the Setup Welcome dialog is displayed
Saturn DigiSizer II 5205
Change
; the Unit IP
Jul 09