Setting up the DigiSizer System

Installing the Ultrasonic Probe

Make sure that no power is supplied to the unit before starting this procedure.
1. Make sure that the power ON/OFF switch on the rear panel of the MasterTech is in the
OFF (O) position.
2. Attach the probe cable to the connector labeled Ultrasonic Probe.
3. Ensure that the probe tip is attached to the probe; if not, attach it. A 1/4-in. probe tip is
supplied with the MasterTech 052. A 1/8-in. tip is also available.
4. Insert the cylinder into the arm, positioning it so that the longer slots slide over the
protrusions inside the arm.
Position the shorter slots of the cylinder over the protrusions if you are using the
1/8-in. probe.
5. Insert the probe into the cylinder.
34
Cable
Saturn DigiSizer II 5205
Ultrasonic Probe
Cylinder
Arm
Jul 09