Setting up the DigiSizer System

Installing the Liquid Sample Handling Unit

A liquid sample handler is required to transfer the sample to the analyzer for analysis. The
sample is transferred through the use of tubing. This tubing is installed on the left side of the
sample handler and inserted through two ports on the right side of the analyzer. Precut tubing
is provided for this purpose; if you are installing the:
Standard unit, the tubing is 1.27-cm (1/2-in.) diameter (2 pieces)
Low-volume unit, the tubing is 0.79-cm (5/16-in.) diameter (1 piece)
Perform the following steps to install the liquid sample handler.
1. Remove the two plugs from the ports on the side of the sample handler
If you are installing the Low-volume unit, the tubing on the upper port is installed; remove
the plug from the end of the tubing.
2. Install the tubing onto the connectors provided on the left side of the liquid sample
handler. One piece of tubing is approximately 2 inches longer than the other; install this
piece on the upper connector and the shorter one on the lower connector.
3. Install the tubing onto the connectors provided on the left side of the liquid sample
handler. One piece of tubing is approximately 2 inches longer than the other; install this
piece on the upper connector and the shorter one on the lower connector.
Longer piece on upper port.
This piece of tubing is
already installed if you have
the Low-volume unit.
Shorter piece on lower port.
14
Saturn DigiSizer II 5205
Jul 09