Verifying Operation

Fire the Laser

This procedure can be performed with or without the analysis cell in place; no water is
required.
1. If the Instrument Schematic is not shown on the computer screen, select Unit[n] > Show
instrument schematic.
2. Select Unit[n] > Enable manual control.
3. Right-click on the Laser-beam Splitter Box.
4. Click Laser On.
5. Observe for ANY response from the Detector.
6. Click Laser Off.

Checking the Temperature

Observe the temperature reading on the schematic. The temperature should come within band
in 15 minutes or less.
70