Saturn DigiSizer II 5205
5. Turn off the unit. You may hear the solenoid valve click as it opens to vent air back into
the filter cartridge.
6. Turn the unit back on. You should again see a fast stream of bubbles, which tapers off after
a few seconds.
Verifying MasterTech Operation
Perform these steps only if the customer purchased a MasterTech.
To verify operation, manually move each subsystem using the manual control option of the
Saturn DigiSizer analysis program.
1. Turn on the MasterTech.
2. Fill a beaker with approximately 80 mL of water.
3. Place the beaker in the number 1 position of the MasterTech tray.
4. To test the ultrasonic probe, remove it from the holder and insert the tip of the probe well
into the water.
NEVER operate the ultrasonic probe without having the tip immersed into a
liquid. Doing so may harm the probe and/or the driving electronics. DO NOT let the tip
of the probe come in contact with a glass beaker. Doing so may shatter the beaker.
5. While holding the probe tip underwater, press the ultrasonic probe AUTO/ON switch. You
will hear a high-pitched sound, indicating that the probe is working properly.
DO NOT press the ultrasonic probe AUTO/ON switch unless the probe and its cable
are properly connected to the MasterTech. Doing so may harm the driving electronics.
6. Press the ultrasonic probe AUTO/ON switch again to turn off the probe.
7. Place the probe back into the holder.
Jul 09
Verifying Operation
77