Installing The Liquid Sample Handler Tubing - Micromeritics Saturn II 5205 DigiSizer Installation Instructions And Checklist

Table of Contents

Advertisement

Saturn DigiSizer II 5205
3. The sample cell is grooved on the left side and angled slightly on the right, front side.
Holding the sample cell by the edges, slide the grooved (left) side of the cell onto the left
vertical bar.
Left grooved side
4. Swing the sample cell into place (against the right vertical bar).
5. Slide the top bracket of the sample cell holder down.
6. Tighten the knurled nuts.

Installing the Liquid Sample Handler Tubing

The following items for installing the tubing are provided in the accessories kit:
Appropriate plastic containers for the analysis, rinse, overflow, and waste (drain) liquids
Precut pieces of Tygon tubing for the analysis, rinse, overflow, and drain function
The Low-volume unit uses the same liquid for analysis and rinsing. Therefore, it does
not require attachment of rinse tubing, rinse pump, or a separate rinse container.
Ignore all references to the rinse tubing and container in this procedure if you have
the Low-volume unit.
.
1. Prepare the containers and connect the tubing as follows:
2. Prepare the containers:
a. Place the rinse and analysis liquid containers on the floor in front of the liquid sample
handler; remove the caps.
b. Pour approximately 10 liters of deiozined water with 0.0005% sodium metaphosphate
(W/W) into the analysis liquid container.
If your rinse and analysis liquids are the same or if you are installing the Low-volume
unit, skip step c. Be sure to access the Unit configuration dialog and specify that the
rinse and analysis liquids are the same.
Jul 09
Right front side
Setting up the DigiSizer System
Knurled nuts
19

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Saturn II 5205 DigiSizer and is the answer not in the manual?

Table of Contents