Verifying Operation

Ultrasonic Probe

1. With manual control enabled, right-click the Ultrasonic Probe symbol on the instrument
schematic screen, then select Probe on.
Ultrasonic
Probe
2. Right-click the Ultrasonic Probe symbol, then select Probe off.
3. Right-click on the Ultrasonic Probe symbol, then select Set probe intensity.
4. The Ultrasonic Intensity Setting dialog is displayed.
a. Enter 40 in the Ultrasonic Intensity field, then repeat steps 1-3.
b. Enter 60 in the Ultrasonic Intensity field, then repeat steps 1-3.
c. Enter 80 in the Ultrasonic Intensity field, then repeat steps 1-3.
d. Enter 100 in the Ultrasonic Intensity field, then repeat steps 1-3.
5. Check the Waste valve:
a. Right-click on the Waste valve displayed in the instrument schematic, then select Open.
b. Right-click on the Waste valve displayed in the instrument schematic, then select Close.
74
Saturn DigiSizer II 5205
Jul 09