Saturn DigiSizer II 5205
Installing the AquaPrep Tubing
The AquaPrep has inlet and outlet connectors onto which tubing is installed for circulating the
water through the AquaPrep for degassing.
1. Install the two pieces of tubing (supplied in the accessories kit) into the connectors located
on the right side of the front panel.
2. Fill a 10-liter, plastic container with deionized water which has been filtered with a 1-
micron filter. Place the container on the floor directly below the AquaPrep.
The AquaPrep works only with water; do not use any other liquid. Do not use water
that contains a detergent or wetting solution.
3. Insert the other ends of the tubing into the water. Orient the tubing as follows:
• The end of the inlet tubing (supply) should be close to the bottom of the container.
• The end of the outlet tubing (return) should be just below (approximately 5 cm, or 2 in.)
the surface of the water so that splashing does not re-introduce air into the liquid.
It is important that the tubing is arranged at different levels. This promotes circulation and
ensures that all the liquid is exposed during the deaeration process.
Jul 09
Tubing Connectors
Setting up the DigiSizer System
27