Verifying Operation
5. Select Report after measurement and choose Screen as the destination.
6. Click
Next;
If this background is unsatisfactory, you can measure another one by clicking Repeat.
Refer to Appendix I of the Saturn DigiSizer 5205 Operator's Manual for guidelines on
determining an appropriate background.
7. Click
Done
if desired.
Creating a Sample Information File
For the analysis test included in this document you will need to create a sample information
file, then start an analysis using the conditions specified in the file. Instructions are included
for standard and confirm systems. Follow the instructions for the type of system purchased by
the customer.
Different lots of reference material may require editing data in these examples.
Standard Systems
Copying the Reference Material File into The Service Directory
A file containing analysis conditions for the reference material test is included on the Saturn
DigiSizer 5205 CD. Using the data in this file will save you time by eliminating the need for
entering analysis conditions individually.
In order to use the provided sample information file, you should first create a Service
subdirectory on the customer's hard drive, then copy the file into the directory as described
below:
1. Create a subdirectory in the C:\5205 directory named Service.
2. Copy this file from the CD into the C:\5205\Service directory:
MRMXXXX.SMP
where XXXX = the serial number of the unit.
82
a progress dialog is displayed, then the background measurement is displayed.
; a report is generated to the screen. You can print the report from this window
Saturn DigiSizer II 5205
Jul 09