Inspecting the Equipment
The documents to be sent to the Micromeritics Service Support Center are:
•
Saturn DigiSizer 5205 Installation Checklist, completed and signed
•
Reference Material Analysis Sample File
•
Laser Diagnostics Report
•
Laser Stability Test Report
•
Calibration files if instrument calibration data are changed during installation.*
*Generally, it is not necessary to change calibration data during installation. If however, you
need to change calibration data, the calibration files must be saved and sent to Micromeritics
for retention in the instrument history file.
Inspecting the Equipment
Before you begin installing the Saturn DigiSizer 5205, check the packing list to ensure that all
components and accessories are available.
If the analyzer has not been unpacked, the customer must unpack the analyzer and
place it on an appropriate work surface before you can proceed with the installation.
If the customer needs to declare equipment as damaged or lost, the customer must
save the shipping cartons. The claims investigator must examine the cartons in order
to complete the inspection report.
The Saturn DigiSizer analyzer weighs 45 kg (99 lbs) and requires the use of two
people to lift it from its shipping carton. One person should not attempt to lift the
analyzer. With one person on each side of the analyzer, lift it upright from its shipping
carton.