Verifying Operation
8. Select Unit [n] > Enable manual control. If the instrument schematic is not displayed,
select Show instrument schematic.
MasterTech Pump
9. Ensure that the Arm AUTO/LOAD switch on the front panel of the MasterTech is in the
AUTO position.
10. Select Unit [n] menu > Initialize MasterTech.
A message is displayed indicating that initialization is in progress. The tray on the
MasterTech rotates in a counterclockwise direction for one beaker position; the arm
lowers to the Rinse position and returns.
The tray then continues to rotate until position 1 is reached. The beaker inserted in the tray
in Step 3 should now be directly under the arm of the MasterTech. Make sure the beaker
contains approximately 80 mL of water.
11. On the instrument schematic, select the symbol representing the MasterTech arm.
When manual mode is enabled, components on the instrument schematic have pop-
up menus displaying the options available for that component. Select the component,
then click the right mouse button to access its pop-up menu.
In each instruction for this step, first access the pop-up menu:
a. Select Move Arm Down two times.
b. Select Ultrasonic On; you will hear a high-pitched sound, indicating that the probe is
operational.
c. Select Ultrasonic Off to turn off the probe.
d. Select Stirrer On; the stirrer begins to operate.
78
MasterTech Arm
MasterTech Arm
Saturn DigiSizer II 5205
Pop-Up Menu
Jul 09