Saturn DigiSizer II 5205
6.
Fuse the input power line according to local safety practices. The input power connector
can be used with either a single-fuse arrangement or a double-fuse arrangement, as shown
in the following illustration.
Fuse Block
a. Observe the position of the fuse block, using the previous figure for reference. If the
single-fuse arrangement is desired, position the fuse block so that the side with the
single-fuse slot and the jumper bar is away from the cover.
If the double-fuse arrangement is desired, position the fuse block so that the side with
the double-fuse slots is away from the cover.
b. If the fuse block is positioned properly for the fusing desired, proceed to Step c.
If the fuse block is not positioned properly for the fusing desired:
1) Remove the fuse block retaining screw.
2) Lift the fuse block from the cover.
3) Rotate the fuse block.
4) Mount the fuse block to the cover.
5) Replace the retaining screw.
The fuses used in the instrument must be identical in type and rating to that
specified. Use of other fuses could result in electrical shock and/or damage to the
instrument.
Jul 09
Retaining Screw
Fuses
Cover
Double-Fuse Arrangement
Fuse
Fuse Block
Jumper Bar
Fuse
Block
Cover
Single-Fuse Arrangement
Appendix C
Fuse
3