Saturn DigiSizer II 5205
Installing the MasterTech (Optional)
The MasterTech is an optional device. Skip these instructions if the customer did not purchase
a Master Tech.
Removing Packing Material From the Unit
After removing the MasterTech from the shipping carton, remove all foam packing material
from the unit before proceeding with installation.
Selecting the Voltage
The MasterTech 052 leaves the factory set for 120 VAC and with the line fuse removed. The
correct setting of the universal power entrance must be checked and the appropriate fuse
installed before the MasterTech can be operated. The MasterTech is designed to operate with
either 100, 120, 220 or 240 VAC at 50 or 60 Hz. Voltage selection and fusing are made at the
power connector, which is located on the rear panel of the unit
On some instruments, the input power connector has a removable cover instead of
the hinged cover described below. If the unit you are installing has a removable cover,
refer to the instructions in Appendix C rather than those below.
1. Make sure the power cord is not connected. The cover of the input power connector
cannot be opened with the power cord installed.
2. Check the voltage setting on the rear panel of the instrument.
• If the voltage is correct:
a. Open the cover by prying the right or left side of the cover open with a pointed object,
then swinging the cover to the left.
Jul 09
Setting up the DigiSizer System
29