Saturn DigiSizer II 5205
Both organic and inorganic particles can be analyzed and measured over a range of 0.06 to
2100 micrometers. The Saturn DigiSizer includes a complete system for circulating the
dispersing liquid/sample mixture through the cell and reservoir, as well as to an external waste
container.

Liquid Sample Handler

The Liquid Sample Handler is available in two configurations:
Standard unit: includes a reservoir that holds 500 to 600 mL of dispersed sample. This
model is best suited for samples containing coarse particles or for those of high density.
Low-Volume unit: includes a reservoir that holds 100 to 120 mL of dispersed sample.
This model is best suited for analyses where the sample quantity or dispersion liquid is
limited, or where the dispersion liquid may be hazardous or difficult to dispose of.
You also can attach a MasterTech 052 which allows you to analyze up to 18 samples
sequentially without further intervention. The MasterTech automatically transfers each sample
to the liquid sample handler for analysis.
Jul 09
The Saturn DigiSizer requires the use of a liquid sample
handler to transfer the sample to the analyzer for analysis.
The sample handler also allows for connection of an
ultrasonic probe to assist in sample dispersion.
The sample handler enables automatic sample handling. All
the user has to do is define analysis conditions (in the
sample file), add the sample to the reservoir (containing the
dispersing liquid), and start the analysis. The sample will be
automatically monitored and diluted when necessary; and
the system rinsed after each analysis (when specified in the
sample file).