Performing a Leak-Check Procedure
1. Turn on the equipment in this order:
a. ultrasonic controller unit
b. MasterTech (if installed) - wait until the MasterTech has finished initialization and
stopped moving before proceeding to Step 2.
c. liquid sample handler
d. analyzer
2. Rinse the system with water to ensure that all tubing and sample cell are installed properly.
3. Open the front panel of the analyzer and remove the cover of the sample cell
compartment.
4. Select Unit > Rinse > DigiSizer from the main menu.
5. Choose the Rinses option; specify 1, then click
6. Observe all tubing connections, as well as the sample cell.
• If leaks occur, click
Drain
4, 5 and 6.
• If leaks do not occur, the DigiSizer is ready to perform analyses.
7. Replace the sample cell compartment cover and close the front panel.