Saturn Digisizer; Performing A Leak-Check Procedure - Micromeritics Saturn II 5205 DigiSizer Installation Instructions And Checklist

Table of Contents

Advertisement

Saturn DigiSizer II 5205

Saturn DigiSizer

Performing a Leak-Check Procedure

1. Turn on the equipment in this order:
a. ultrasonic controller unit
b. MasterTech (if installed) - wait until the MasterTech has finished initialization and
stopped moving before proceeding to Step 2.
c. liquid sample handler
d. analyzer
2. Rinse the system with water to ensure that all tubing and sample cell are installed properly.
3. Open the front panel of the analyzer and remove the cover of the sample cell
compartment.
4. Select Unit > Rinse > DigiSizer from the main menu.
5. Choose the Rinses option; specify 1, then click
6. Observe all tubing connections, as well as the sample cell.
• If leaks occur, click
Drain
4, 5 and 6.
• If leaks do not occur, the DigiSizer is ready to perform analyses.
7. Replace the sample cell compartment cover and close the front panel.
Jul 09
Cancel
to stop the rinsing function. Close the front panel and select
on the Unit menu to drain the sample cell. Correct the problem, and repeat steps
Start
.
Verifying Operation
67

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Saturn II 5205 DigiSizer and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents