Setting up the DigiSizer System
Connecting the Ultrasonic Controller Unit
Connect the ultrasonic controller unit to the probe as follows:
1. Place the ultrasonic controller unit on top of the liquid sample handler.
2. Connect one end of the 9-pin RS232 M/F straight communication cable to the I/O
connector on the rear of the controller unit.
3. Insert one end of the ultrasonic probe BNC cable into the probe connector.
4. Insert appropriate fuse(s) for the input power source. Refer to the following chart for the
appropriate fuse rating.
The fuses used in the ultrasonic controller unit must be identical in type and rating to
that specified. Use of other fuses could result in electrical shock and/or damage to
the unit
Power Source
100-120 VAC
200-240 VAC
5. Ensure that the line voltage is set to match that of the input power source. Refer to the
manufacturer's User's Guide shipped with the ultrasonic probe for instructions on setting
the line voltage.
Do not connect the ultrasonic power cord to a power source until the proper voltage
selection is made. Doing so could result in electrical shock and/or damage to the
ultrasonic unit.
6. Connect the power cord to the power connector and plug into an appropriate power
source.
7. On the rear panel of the liquid sample handler in the area labeled Ultrasonic Probe:
a. Connect the other end of the communication cable (from the ultrasonic controller) to the
connector labeled Controller.
22
I/O Connector
Probe Connector
(for communication
cable)
Fuse
3AG, 3.0 Amp Slow-Blow
5x20 mm, 1.6 Amp Slow-Blow (Type T)
Saturn DigiSizer II 5205
Voltage Selector Card
Power Switch
Power Connector
Jul 09