Saturn DigiSizer II 5205
Low-Volume Liquid Sample Handler
1. Locate the T-fitting of the stainless steel tubing of the drain manifold.
2. Use the 7/16- and the 1/2-in. wrenches provided in the accessories kit for this step. Hold
the T-fitting secure with the 7/16-in. wrench, then use the 1/2-in. wrench to loosen and
remove the topmost nut.
3. Remove the stainless steel plug from the nut. Be sure to store the plug in a secure location.
If the MasterTech is disconnected, this plug will need to be reinstalled in the liquid sample
handler.
4. Insert the stainless steel connector into the nut (the barbed end should be on the outside).
5. Reattach the nut to the T-fitting; tighten finger-tight. Then, using the two wrenches as
shown above, tighten the nut 1/4 turn.
6. Insert one end of the flexible tubing (provided in the accessories kit) through the hole
provided on the base of the sample handler.
Jul 09
T-fitting
Setting up the DigiSizer System
41