Interior Components; Tubing Slots And Tubing Chase; Leak Sensor And Leak Drain - Thermo Scientific Dionex ICS Series Operator's Manual

Vwd variable wavelength detector
Hide thumbs Also See for Dionex ICS Series:
Table of Contents

Advertisement

2.2

Interior Components

To access the interior components of the Dionex VWD, grasp the front cover by
the sides and pull it straight off.
2
1
2
Figure 2-3. Dionex VWD Interior View (Front Cover Removed)
2.2.1

Tubing Slots and Tubing Chase

The tubing slots route tubing (for example, the flow cell inlet tubing)
from the Dionex VWD interior to outside the detector.
The tubing chase routes tubing (for example, the waste line from the flow
cell outlet) from the interior of the Dionex VWD, through the body of the
module, and to the rear panel (see
2.2.2

Leak Sensor and Leak Drain

A leak sensor is installed inside the detector (see
sensor detects a leak, the
error message appears in the Chromeleon Audit Trail. The drip tray
collects the fluid and the leak drain routes the fluid from the drip tray to
the drain port on the Dionex VWD rear panel (see
Doc. 065377-01 1/12
1
3
1
Tubing Slots
Tubing Chase
3
Leak Sensor
ALARM
1
4
5
6
4
Flow Cell
Lamp Housing Cover
5
6
Leak Drain
Figure
2-5).
Figure
LED on the status bar lights, and an
Figure
2 • Description
1
2-3). If the leak
2-5). After
11

Advertisement

Table of Contents
loading

Table of Contents