Thermo Scientific Sorvall BP 8 Instruction Manual
Hide thumbs Also See for Sorvall BP 8:
Table of Contents

Advertisement

Thermo Scientific
Sorvall BP 8 / 16,
Sorvall BIOS 16
Centrifuge and Heavy Duty Centrifuge
Instruction Manual
50149590-l • 2022-02
Visit us online to register your warranty:
thermofisher.com/labwarranty

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Sorvall BP 8 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Thermo Scientific Sorvall BP 8

  • Page 1 Thermo Scientific Sorvall BP 8 / 16, Sorvall BIOS 16 Centrifuge and Heavy Duty Centrifuge Instruction Manual 50149590-l • 2022-02 Visit us online to register your warranty: thermofisher.com/labwarranty...
  • Page 2: Table Of Contents

    Items Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii Intended Use for Sorvall BP 8 / 16 and Sorvall BP 8 / 16 Heavy Duty . . . . . . . . x Medical ..........................x...
  • Page 3 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges 1 . Switch on the Centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .III–1 2 .
  • Page 4 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges 8 . Shipping and Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV–9 Troubleshooting .
  • Page 5 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges HAEMAFlex 12 Rotor / 6 x 2000 mL Swinging Bucket Rotor . . . . . . . . . . . . . B–11 6 x 2000 mL Swinging Bucket Rotor .
  • Page 6 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Table of Figures Figure I–1: Centrifuge clearance, example with barcode scanner .......... I–2 Figure I–2: Position of the bubble in the water level ............. I–4 Figure I–3: Mains connection on the backside of the centrifuge ..........
  • Page 7 Table I–1: Overview of necessary tools for setting up ............I–3 Table V–1: Error messages ....................V–3 Table A–1: Technical Data Thermo Scientific Sorvall BP 8 / 16 / BIOS 16 and Heavy Duty ..A–1 Table A–2: Directives, Standards and Guidelines Sorvall BIOS 16 / Sorvall BIOS 16 Heavy Duty .........................A–2...
  • Page 8: Preface

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Preface Preface Before starting to use the centrifuge, read through this instruction manual carefully and follow the instructions. Failure to follow the instructions and safety information in this instruction manual will result in the expiration of the sellers Warranty.
  • Page 9: Table Preface-1: Items Supplied

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Preface Article No . Item Quantity 20280119 Bubble Level Instruction Manual Table Preface–1: Items Supplied If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.
  • Page 10: Intended Use For Sorvall Bp 8 / 16 And Sorvall Bp 8 / 16 Heavy Duty

    6 and HAEMAFlex 8 rotor. The Thermo Scientific Sorvall BP 16 and Thermo Scientific Sorvall ™ ™ BP 16 Heavy Duty centrifuge can be operated with the Thermo Scientific HAEMAFlex 12 and HAEMAFlex 16 rotor. For more details on buckets and further components: [à  2 A–6] [à ...
  • Page 11: Signal Words And Colors

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Preface Signal Words and Colors Signal Word and Color Degree of Hazard WARNI NG Indicates a hazardous situation that, if not avoided, could result in death or serious injury.
  • Page 12 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Preface Risk from handling hazardous substances. Especially when working with corrosive samples (salt solutions, acids, bases), the components and the centrifuge chamber have to be cleaned thoroughly.
  • Page 13 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Preface Physical harm caused by ignoring operative basics. Never use the centrifuge if parts of its casing is damaged or missing. Never start the centrifuge when the centrifuge door is open.
  • Page 14: Symbols Used On The Centrifuge And Its Components

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Preface Symbols used on the Centrifuge and its Components This symbol refers to general hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe.
  • Page 15 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Preface Indicates the date after which the device is not to be used. Indicates the manufacturer’s batch code so that the batch or lot can be identified.
  • Page 16: Symbols Used In The Manual

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Preface Indicates conformity with Underwriter Laboratories (UL) requirements. Indicates the maximum possible weight of the adapter fully loaded. Indicates the maximum rcf value for centrifugation.
  • Page 17: Transport And Set Up

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Transport and Set Up THE ACTIVITIES DESCRIBED IN THIS CHAPTER ARE FOR SERVICE TECHNICIANS. Transport and Set Up CA UT I O N Make sure that the centrifuge and the packaging is not damaged. Speak to the shipping company and Thermo Fisher Scientific immediately if damage is found.
  • Page 18: Transporting

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Transport and Set Up THE ACTIVITIES DESCRIBED IN THIS CHAPTER ARE FOR SERVICE TECHNICIANS. Figure I–1: Centrifuge clearance, example with barcode scanner 2. Transporting WA R N I N G Never stand before a moving centrifuge to stop it.
  • Page 19: Setting Up

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Transport and Set Up THE ACTIVITIES DESCRIBED IN THIS CHAPTER ARE FOR SERVICE TECHNICIANS. 3. Setting Up CA UT I O N Do not push at the touchscreen.
  • Page 20: Level The Centrifuge

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Transport and Set Up THE ACTIVITIES DESCRIBED IN THIS CHAPTER ARE FOR SERVICE TECHNICIANS. Level the Centrifuge CA UT I O N If you do not level the centrifuge, the centrifuge can crash because of imbalance.
  • Page 21: Mains Connection

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Transport and Set Up THE ACTIVITIES DESCRIBED IN THIS CHAPTER ARE FOR SERVICE TECHNICIANS. 4. Mains Connection N OT I C E Connect the centrifuge into grounded electrical sockets only.
  • Page 22: Shipping

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Transport and Set Up THE ACTIVITIES DESCRIBED IN THIS CHAPTER ARE FOR SERVICE TECHNICIANS. 6. Shipping WA R N I N G Before shipping the centrifuge and components you must clean and if necessary disinfect or decontaminate the full system.
  • Page 23: Thermo Scientific

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Thermo ScientificCentri-Touch User Interface Thermo Scientific Centri-Touch User Interface Actual status information ➀ Set values ➁ Settings and navigation ➂  II–1...
  • Page 24: Status

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Thermo ScientificCentri-Touch User Interface Status The upper part of the touchscreen displays the centrifuge status. The remaining time is indicated while centrifugation is in progress. The progress bar shows the actual phase of the centrifugation.
  • Page 25: Run Parameters

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Thermo ScientificCentri-Touch User Interface Run Parameters The set points for centrifugation can be set in the parameterization window. Clicking one of these buttons will open a new window in which you can input the appropriate set point value.
  • Page 26: Lighthouse Mode

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Thermo ScientificCentri-Touch User Interface Lighthouse Mode If no entry is made via the main screen in 30 seconds, the unit changes to lighthouse mode. In this mode, the status, current values and progress display are displayed larger.
  • Page 27: Operation

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation I I I Operation For detailed instructions and information refer to the separate manual for the Thermo Scientific Centri-Touch User Interface. ™ ™...
  • Page 28: Before Installation

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation 3. 1. Before Installation Make sure that all rotor parts are clean, dry and have no nicks and scratches. Always use a rotor body, a windshield bowl and a windshield lid that have the same article number.
  • Page 29: Rotor Installation

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation 3. 2. Rotor Installation CA U T I O N An an improper installation of the rotor can cause severe damage to the rotor and the centrifuge.
  • Page 30: Figure Iii-4: Rotor Locking Tool

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation the rotor body carefully on the drive shaft. Turn the rotor body against the windshield bowl until the locating pin engages into the rotor body.
  • Page 31: Figure Iii-6: Tighten The Rotor To The Drive Shaft

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation Figure III–6: Tighten the rotor to the drive shaft. Clean and lubricate the rotor trunnion bolts and the two bolt seats on the sides of each bucket. [à  2 IV–3]...
  • Page 32: Figure Iii-9: Correctly Assembled Round Bucket (75003836/75003936, Not 75003964)

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation Figure III–8: Install buckets into rotor, e.g. HAEMAFlex 12 When assembling round buckets (75003836/75003936) insert the O-ring below the thread in the bucket. Do not insert it to the bucket lid.
  • Page 33: Figure Iii-10: Correct Loading

    Operation Correct Loading ✓ Figure III–10: Correct Loading Incorrect Loading ✕ Figure III–11: Incorrect Loading Put the windshield lid into the holder at the inner side of the centrifuge door. The Thermo Scientific Auto-Lid function ™ ™ automatically opens and closes the windshield when the centrifuge door opens and closes.
  • Page 34: Maximum Loading

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation Figure III–12: Insert windshield lid into holder CA U T I O N Always put the windshield lid into the holder at the inner side of the centrifuge door. Make sure the rotor is locked properly.
  • Page 35: Use Of Tubes And Consumables

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation CA U T I O N If the maximum compartment mass is exceeded, the speed or fill level must be reduced. Failure to do so can result in personal injury and / or centrifuge damage.
  • Page 36: Preselect Speed / Rcf-Value

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation 6. 2. Preselect Speed / RCF-Value Press on the speed field on the main screen. Press rpm or RCF. The selected function is highlighted in yellow.
  • Page 37: Pre-Warm Or Pre-Cool The Centrifuge

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation 6. 5. Pre-warm or Pre-cool the Centrifuge To pre-temp the centrifuge, proceed as follows: Use Configuration and Runs to select a stored program. The pre-temp function is connected to a program.
  • Page 38: Centrifugation

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation 8. Centrifugation WA R N I N G Do not operate the centrifuge on its casters. CA U T I O N An an improper installation of the rotor can cause severe damage to the rotor and the centrifuge.
  • Page 39: Using The Drain Box

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation 8. 3. Using the Drain Box You can use the drain box (75007730) to collect water from of the centrifugation chamber. Figure III–14: Drain box (75007730) Install the drain box.
  • Page 40: Remove A Rotor

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation N OT I C E Don’t use any sharp tool to extract the plug. This might damage the centrifugation chamber. You have to regularly check if the drain box is filled and empty the drain box to avoid spilling on the floor.
  • Page 41: Switch Off The Centrifuge

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Operation CA U T I O N The centrifuge drive can be damaged if the rotor body falls on it. Hold the windshield knob with one hand and at the rim of the windshield bowl with the other.
  • Page 42: Maintenance And Care

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Maintenance and Care Maintenance and Care 1. Cleaning Intervals For the sake of personal, environmental and material protection, you must clean and if necessary, disinfect the centrifuge and its components on a regular basis.
  • Page 43: Rotor And Components Inspection

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Maintenance and Care 2. 1. Rotor and Components Inspection CA UT I ON Do not run any rotor or components with sign of damage.
  • Page 44: Cleaning

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Maintenance and Care For detailed instructions and information refer to the separate manual for the Thermo Scientific CentriTouch User Interface. 3. Cleaning CAU T I ON Before using any cleaning methods, users should check with the manufacturer of the cleaning agents that the proposed method will not damage the equipment.
  • Page 45: Touchscreen

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Maintenance and Care Clean the bucket surfaces and rotor bolts as follows: Cut a cleaning pad (66309) into 12 mm (1/2 inch) wide strips.
  • Page 46: Filter Mat

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Maintenance and Care 3. 2. Filter Mat CA UT I O N Do not touch the edges of the housing with your hands when the grid is removed.
  • Page 47: Disinfection

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Maintenance and Care 4. Disinfection WA R N I N G Do not touch infected parts. Hazardous infection is possible when touching the contaminated rotor and centrifuge parts. Infectious material can get into the centrifuge when a tube breaks or as a result of spills.
  • Page 48: Autoclaving

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Maintenance and Care 6. Autoclaving CA UT I O N Never exceed the permitted temperature and duration when autoclaving. N OT I C E No chemical additives are permitted in the steam.
  • Page 49: Maintenance

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Maintenance and Care 7. Maintenance 7. 1. Preventive Maintenance In order to keep this product able to perform the intended application in a reliable and safe state, ongoing preventive maintenance is necessary in accordance with the following recommended schedule: For rotors and buckets mind the information in the Rotor and Components Inspection chapter.
  • Page 50: Shipping And Disposal

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Maintenance and Care 8. Shipping and Disposal WA R N I N G When removing the centrifuge and components from use for disposal you have to clean and if necessary disinfect or decontaminate the entire system.
  • Page 51: Troubleshooting

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Troubleshooting Troubleshooting 1. Mechanical Emergency Door Release WA R N I N G Serious injuries can occur, if you touch a spinning rotor with your hands or tools.
  • Page 52: Ice Formation

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Troubleshooting Pull out the power supply plug. On both sides of the housing are two white plastic plugs. Get these two white plastic plugs out of both sides of the housing.
  • Page 53: Troubleshooting By Guide

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Troubleshooting 3. Troubleshooting by Guide CAU T I ON If problems occur other than those listed in this table, service technician must be contacted.
  • Page 54: Information For The Customer Service

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Troubleshooting 4. Information for the Customer Service If you need to contact a service technician, please provide the order no. and the serial no. of your centrifuge. This information can be found on the back near the inlet for the power supply cable.
  • Page 55: Technical Specifications

    1 m in front of the instrument at 1.6 m height. Without rotor. ³ Applications using temperatures below 25 °C must pre-cool the centrifugation chamber and rotor. Table A–1: Technical Data Thermo Scientific Sorvall BP 8 / 16 / BIOS 16 and Heavy Duty  A–1...
  • Page 56: Directives, Standards And Guidelines

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Technical Specifications 1. Directives, Standards and Guidelines Centrifuge Region Directive Standard Thermo Scientific Europe 2006/42/EC Machinery EN 61010-1 Sorvall BIOS 16 Single Phase 2014/35/EU Low Voltage...
  • Page 57: Table A-3: Directives, Standards And Guidelines Sorvall Bp 8 / 16 And Heavy Duty

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Technical Specifications Centrifuge Region Directive Standard Thermo Scientific Europe 93/42/EEC Medical Devices EN 61010-1 Sorvall BP 8 Single Phase 98/79/EC In Vitro Diagnostics IEC 61010-2-020...
  • Page 58: Refrigerants

    1.80 kg 15 bar 34 bar 1 397 2.51 t Sorvall BIOS 16 Contains fluorinated greenhouse gases in a hermetically sealed system. Table A–4: Refrigerants used in the Sorvall BP 8 / 16 / BIOS 16 and Heavy Duty  A–4...
  • Page 59: Mains Supply

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Technical Specifications 3. Mains Supply The following data is to be taken into consideration, when selecting the mains connection socket. Power Mains Voltage Rated Current...
  • Page 60: Components

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Technical Specifications 4. Components 4. 1. Sorvall BP 8 / 16 and Heavy Duty Autoclavable Max . (121 °C, 20 min; Article No .
  • Page 61: Table A-6: Components For Sorvall Bp 8 / 16 And Heavy Duty

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Technical Specifications Autoclavable Max . (121 °C, 20 min; Article No . Description Rotor Capacity Max . RCF Speed = no, = yes ✘...
  • Page 62: Sorvall Bios 16 And Heavy Duty

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Technical Specifications 4. 2. Sorvall BIOS 16 and Heavy Duty Autoclavable Max . (121 °C, 20 min; Article No . Description Rotor Capacity Max . RCF...
  • Page 63: Table A-7: Components For Sorvall Bios 16 And Heavy Duty

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Technical Specifications Autoclavable Max . (121 °C, 20 min; Article No . Description Rotor Capacity Max . RCF Speed = no, = yes ✘...
  • Page 64: Specifications Of Centripak Bpc Adapter

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Technical Specifications Do not use rotors, buckets or components which show any signs of removed protective coating, WA R N I N G corrosion or cracks.
  • Page 65: Rotors

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors Rotors Items Supplied for Rotors Article No . Item Quantity Rotor (incl . Windshield Bowl and Windshield Lid) 75003861 Thermo Scientific HAEMAFlex 6...
  • Page 66: Haemaflex 6 Rotor

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 6 Rotor with Single Blood Bag Buckets (oval) Technical Data Sorvall BP 8 / Sorvall BP 8 Heavy Duty 200, 208, 220,...
  • Page 67: Table B-4: Haemaflex 6 Rotor With Single Blood Bag Buckets (Round

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 6 Rotor with Single Blood Bag Buckets (round) Technical Data Sorvall BP 8 / Sorvall BP 8 Heavy Duty 200, 208, 220,...
  • Page 68: Haemaflex 6 Rotor

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 6 Rotor with Single Blood Bag Buckets with Filter Pack Technical Data Sorvall BP 8 / Sorvall BP 8 Heavy Duty 200, 208, 220,...
  • Page 69: X 1000 Ml Swinging Bucket Rotor

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors 6 x 1000 mL Swinging Bucket Rotor with Round Buckets Technical Data Sorvall BIOS 16 / Sorvall BIOS 16 Heavy Duty Centrifuge Voltage...
  • Page 70: Haemaflex 8 Rotor

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 8 Rotor with Single Blood Bag Buckets (oval) Technical Data Sorvall BP 8 / Sorvall BP 8 Heavy Duty 200, 208, 220,...
  • Page 71: Table B-8: Haemaflex 8 Rotor With Single Blood Bag Buckets (Round

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 8 Rotor with Single Blood Bag Buckets (round) Technical Data Sorvall BP 8 / Sorvall BP 8 Heavy Duty 200, 208, 220,...
  • Page 72: Haemaflex 8 Rotor

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 8 Rotor with Single Blood Bag Buckets with Filter Pack Technical Data Sorvall BP 8 / Sorvall BP 8 Heavy Duty 200, 208, 220,...
  • Page 73: X 1000 Ml Swinging Bucket Rotor

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors 8 x 1000 mL Swinging Bucket Rotor with Round Buckets Technical Data Sorvall BIOS 16 / Sorvall BIOS 16 Heavy Duty Centrifuge Voltage...
  • Page 74: Haemaflex 12 Rotor / 6 X 2000 Ml Swinging Bucket Rotor

    Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 12 Rotor / 6 x 2000 ml Swinging Bucket Rotor with Double Blood Bag Buckets Technical Data Sorvall BP 16 / Sorvall BP 16 Heavy Duty / Sorvall BIOS 16 / Sorvall BIOS 16 Heavy Duty...
  • Page 75 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 12 Rotor / 6 x 2000 mL Swinging Bucket Rotor with Oval Buckets Technical Data Sorvall BP 16 / Sorvall BP 16 Heavy Duty / Sorvall BIOS 16 / Sorvall BIOS 16 Heavy Duty...
  • Page 76 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors 6 x 2000 mL Swinging Bucket Rotor with Round Buckets Technical Data Sorvall BIOS 16 / Sorvall BIOS 16 Heavy Duty Centrifuge Voltage...
  • Page 77 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 16 Rotor / 8 x 2000 ml Swinging Bucket Rotor with Double Blood Bag Buckets Technical Data Sorvall BP 16 / Sorvall BP 16 Heavy Duty / Sorvall BIOS 16 / Sorvall BIOS 16 Heavy Duty...
  • Page 78 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors HAEMAFlex 16 Rotor / 8 x 2000 mL Swinging Bucket Rotor with Oval Buckets Technical Data Sorvall BP 16 / Sorvall BP 16 Heavy Duty / Sorvall BIOS 16 / Sorvall BIOS 16 Heavy Duty...
  • Page 79 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Rotors 8 x 2000 mL Swinging Bucket Rotor with Round Buckets Technical Data Sorvall BIOS 16 / Sorvall BIOS 16 Heavy Duty Centrifuge Voltage...
  • Page 80 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–1...
  • Page 81 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–2...
  • Page 82 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–3...
  • Page 83 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–4...
  • Page 84 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–5...
  • Page 85 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–6...
  • Page 86 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–7...
  • Page 87 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–8...
  • Page 88 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–9...
  • Page 89 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Chemical Compatibility Chart  C–10...
  • Page 90 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges Index Ice Formation V–2 Install a Rotor III–1 Intended Use Intended User Items Supplied viii Acceleration / Deceleration Profiles III–9 Items Supplied for Rotors B–1...
  • Page 91 Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges with Oval Buckets B–11 8 x 1000 mL Swinging Bucket Rotor with Round Buckets B–9 8 x 2000 mL Swinging Bucket Rotor with Oval Buckets B–14...
  • Page 92 Thermo Scientific Sorvall BP 8 / 16 Thermo Scientific Sorvall BIOS 16 Thermo Scientific Sorvall BP 8 / 16 Heavy Duty Thermo Scientific Sorvall BIOS 16 Heavy Duty 1434 Thermo Electron LED GmbH Zweigniederlassung Osterode Am Kalkberg, 37520 Osterode am Harz Germany thermofisher.com/centrifuge...