Replacing The Mercury Burner - Olympus MX63 Instructions Manual

Semiconductor/fpd/industrial inspection microscopes
Table of Contents

Advertisement

6-4 Replacing the mercury burner

a
70 70
Applicable lamp
1
4
2
a
2
b
USH-103OL (made by Ushio Co.)
Removing the mercury lamp housing
1
Set the main switch of the power supply for mercury burner (U-RFL-T)
to
(OFF) and disconnect the power cord.
2
Remove the connector of the lamp housing from the power supply
(U-RFL-T).
3
Loosen the mounting screw a of the adapter (MX-HGAD) using the
Allen screwdriver.
Be careful, if the mounting screw is loosened too much, it
NOTE
may come off.
4
Remove the lamp housing.
For details of the power supply for mercury burner
(U-RFL-T), refer to the instruction manual provided with
the unit.
Replacing the mercury burner
1
Loosen the mounting screw a on the top of the lamp housing using
the Allen screwdriver.
2
Hold the upper section of the lamp housing and pull it upward to
remove the socket part.
To prevent malfunctions, do not hold the lamp housing by
NOTE
the centering knobs b .